There has been a significant resurgence in the development of antibody-drug conjugates (ADC) as target-directed therapeutic agents for cancer treatment. Among the factors critical to effective ADC design is the Drug Antibody Ratio (DAR). The DAR describes the degree of drug addition that directly impacts both potency and potential toxicity of the therapeutic, and can have significant effects on properties such as stability and aggregation. Determination of DAR is, therefore, of critical importance in the development of novel ADC therapeutics.
DAR is typically assessed by mass spectrometry (MALDI–TOF or ESI–MS) or UV spectroscopy. Calculations based on UV absorption are often complicated by similarities in extinction coefficients of the antibody and small molecule. Mass spectrometry, though a powerful tool for Mw determination, depends on uniform ionization and recovery between compounds - which is not always the case for ADCs.
Here we present a method for DAR determination based on
size-exclusion chromatography combined with multi-angle light scattering (SEC–MALS) in conjunction with UV absorption and differential refractive index detection. Figure 1 shows UV traces for two model ADCs; molecular weights of the entire ADC complexes are determined directly from light scattering data.
Figure 1: Molar masses for two distinct ADC formulations are determined using SEC–MALS analysis.
Component analysis is automated within the ASTRA 6 software package by using the differential refractive index increments (dn/dc) and extinction coefficients, which are empirically determined for each species or mined from the literature, to calculate the molar mass of the entire complex as well as for each component of the complex.
Figure 2: Molar masses for the antibody and total appended drug are calculated in the ASTRA software package based on prior knowledge of each component’s extinction coefficent and dn/dc, allowing determination of DAR based on a nominal Mw of 1250 Da for an individual drug.
In this example an antibody has been alkylated with a compound having a nominal molecular weight of 1250 Da (Figure 2). Molar masses of the antibody fractions are similar, which indicates that the overall differences between the two formulations reflect distinct average DARs that are consistent with values obtained by orthogonal techniques. Note that the molar mass traces for the conjugated moiety represent the total amount of attached pendant groups; the horizontal trends indicate that modification is uniform throughout the population eluting in that peak.
Wyatt Technology Corporation
6300 Hollister Avenue, Santa Barbara, California 93117, USA
Tel: +1 (805) 681 9009 fax: +1 (805) 681 0123
Website: www.wyatt.com
Inside the Laboratory: The Gionfriddo Group at the University at Buffalo
March 28th 2024In this edition of “Inside the Laboratory,” Emanuela Gionfriddo, PhD, an associate professor of chemistry at the University at Buffalo, discusses her group’s current research endeavors, including using solid-phase microextraction (SPME) coupled to liquid chromatography (LC) and gas chromatography (GC) to further understand the chemical relationship between environmental exposure and disease and elucidate micropollutants fate in the environment and biological systems.
Transferring Methods to Compact and Portable HPLC
February 14th 2024The current trend in laboratory equipment design is the miniaturization of laboratory instruments. Smaller-scale HPLC instruments offer benefits that cannot be matched by analytical-scale equipment, especially in the areas of portability, reduced fluid volumes, and reduced operating costs. Yet, the miniaturization of laboratory equipment has brought with it a unique set of challenges, including transferring methods to compact LC. Capillary LC expands the use of LC to applications not currently done using conventional LC in a wide array of application areas, including pharmaceutical, food and beverage, petrochemical, environmental, and oil and gas. Greg Ward, Axcend’s CEO wrote, “Customers want an HPLC system with a small footprint, low flow rates and green chemistry.” Join his podcast where he shares method transfer in these application areas.