From Microwave to Chromatograph: A Faster, Greener Route to Food Lipid Analysis

Listen
0:00 / 0:00

Key Takeaways

  • MTBE-based microwave-assisted extraction achieved <10% statistical differences versus legacy methods, with solvent choice influencing FAME recovery more than extraction technique.
  • GC×GC-TOFMS was reserved for identity confirmation and interferent checks, enabling efficient GC-FID quantitation across high replicate loads without compromising comparability.
SHOW MORE

LCGC International spoke to Giulia Giacoppo and Marco Beccaria from the University of Ferrara about how microwave-assisted MTBE extraction, combined with solid-phase fractionation, GC-FID, and UHPLC-APCI-MS, matches gold-standard lipid extraction methods and reveals the detailed fat profile of Sicily's Bronte pistachio.

Getting fat out of food so it can be properly analyzed has traditionally been a slow, solvent-heavy process. Two recent studies suggest there may be a better way. Using microwaves together with the solvent MTBE (methyl tert-butyl ether), the research team responsible for both studies first put this faster approach head-to-head with the long-established "gold standard" extraction methods, using one of Italy's most celebrated products, the Bronte pistachio from Sicily, as a case study. They found it delivered nearly identical fatty acid profiles among the tested extraction methods, with few statistical differences (under 10%). They also weighed how environmentally friendly each option was. They then took the technique beyond the extraction method comparison and applied it to a deeper investigation of the intact lipids present in the same nut. There they built a complete analytical workflow around it, sorting the extracted fat into its main families before analysis.

Chromatography, the science of separating a mixture into its individual components, lies at the heart of both projects. Gas chromatography (GC) was used to map fatty acids in both studies, and liquid chromatography paired with mass spectrometry was used to identify dozens of whole fat molecules in the pistachios.

LCGC International spoke to Giulia Giacoppo and Marco Beccaria of Italy’s University of Ferrara, two of the authors of the pair of papers which stemmed from these studies,1,2 to talk about what these findings mean for food analysis, why chromatography remains such a powerful tool, and how a faster, greener extraction method could change the way scientists study what's in our food. Moreover, the authors claimed that this analytical workflow can be easily adapted to other fields where lipids are the target molecules (e.g. clinical chemistry and petrochemistry).

Why did you choose to pair two-dimensional gas chromatography with time-of-flight mass spectrometry (GC×GC-TOFMS) for qualitative profiling alongside a separate GC-FID system for quantification, rather than relying on a single chromatographic setup for both tasks?

Giulia Giacoppo (GG): The two platforms were chosen because they answer two different questions. The GC×GC-TOFMS run was used qualitatively, to confirm peak identity and to verify that no co-extracted interferents were present under the peaks of interest. This analysis revealed approximately fifteen fatty acid methyl esters (FAMEs), but nearly half were present only at trace levels and therefore had negligible impact on the comparison of extraction methods. Since the objective of the study was to compare FAME profiles rather than perform an exhaustive compositional characterization, GC-flame ionization detector (FID) was considered fully adequate for quantification. Once peak identities had been confirmed by GC×GC-TOFMS, the simpler GC-FID method provided all the information required for a robust comparison of samples analyzed under identical conditions. Given the large analytical workload (seven extraction methods, extraction and GC replicates, blanks, and FAME standards), restricting GC×GC-TOFMS to identity confirmation while using GC-FID for routine quantification represented the most appropriate and proportionate analytical strategy.

What specific separation advantages does GC×GC offer over conventional one-dimensional GC when resolving complex FAME profiles in a matrix as lipid-rich as pistachio oil?

Marco Beccaria (MB): It is well established that GC×GC provides enhanced sensitivity and substantially higher peak capacity than conventional one-dimensional GC, enabling the generation of structured group-type patterns and a more comprehensive characterization of FAMEs in lipid matrices. Although pistachio oil, like olive oil and other vegetable oils, is generally considered a relatively simple lipid matrix in terms of fatty acid (FA) composition, GC×GC remains particularly valuable for detecting minor constituents, including odd-chain FAs, isomers, and for resolving FAs that may co-elute under one-dimensional GC conditions.

How did the choice of extraction solvent and method— methyl-tert-butyl ether (MTBE)-based microwave-assisted extraction (MAE) versus Soxhlet with n-hexane, MTBE, or diethyl ether—influence the resulting chromatographic FAME profiles, and did you observe any solvent-dependent shifts in peak resolution or compound recovery?

GG: Since the derivatization process and chromatographic conditions were identical for all the samples, the main differences concerned the relative abundance of individual FAMEs after the different extractions. The MAE methods generally yielded slightly higher proportions of polyunsaturated fatty acids (PUFAs) compared with Soxhlet extractions, whereas Soxhlet extracts tended to be relatively enriched in saturated fatty acids (SFAs) and monounsaturated fatty acids (MUFAs). In addition, the biphasic MTBE/MeOH/H₂O procedures (Matyash and simplified Matyash) produced FAME profiles that differed from those obtained with the single-solvent methods, showing relatively higher palmitic and linoleic acid contents and lower oleic acid contents. These differences suggest that extraction conditions may influence the relative recovery of lipid fractions with distinct FA compositions. This behavior was also reflected in the PCA, where the single-solvent methods clustered closely together, while the Matyash-based methods showed greater dispersion. The close similarity between Soxhlet-MTBE and MAE-MTBE further indicates that solvent selection may have had a stronger influence on the resulting FAME profile than the extraction technique itself.

Since FAMEs require derivatization before GC analysis, did the extraction method itself introduce any variability into the derivatization step or downstream chromatographic separation that needed to be controlled for?

GG: Any potential effect of the extraction solvent on derivatization or chromatographic separation was controlled by evaporating all extracts to dryness, applying the same derivatization protocol, and reconstituting the resulting FAMEs in the same injection solvent prior to GC analysis. These measures were implemented to minimize any variability arising from sample preparation and analysis, ensuring that the observed differences in FAME profiles reflected the extraction methods themselves rather than downstream analytical artefacts.

When comparing FAME composition data across the different extraction techniques via one-way analysis of variance (ANOVA) and a post-hoc comparisons (Tukey's Honestly Significant Difference [HSD]), were there particular fatty acid peaks or regions of the chromatogram that showed the most sensitivity to the extraction method used?

MB: The effect of the extraction method was not evenly distributed across the chromatogram. Most FAMEs showed relatively stable abundances across extraction procedures, whereas significant differences were concentrated in a limited number of fatty acids. Palmitic acid (P, C16:0) was the most discriminating FAME and was primarily responsible for distinguishing the Matyash-based microwave methods from the Soxhlet extractions in the ANOVA/Tukey analysis. Smaller but still significant differences were observed in the C18-C20 region, particularly for oleic (O, C18:1), linoleic (L, C18:2), linolenic (Ln, C18:3), and arachidic (A, C20:0) acids. Overall, the statistical differences were therefore confined to a small subset of FAMEs rather than being distributed across the entire chromatographic profile, indicating that the extraction methods had a selective impact on specific FAMEs while leaving most of the profile unchanged.

How did operating the atmospheric pressure chemical ionization (APCI) interface in both positive and negative polarity modes expand your chromatographic coverage of triacylglycerols (TAGs) versus phospholipids—did each polarity favor detection of different lipid classes or ionization behaviors?

GG: TAGs’ fraction was analyzed in APCI positive mode alone, since protonated molecular ions and their diacylglycerol-type fragments, together with GC information, already provide the information needed for assignment and, when possible, for locating the sn-2 acyl chain.

Phospholipids' (PLs) fractions were instead run in both polarities. In APCI(+), since phosphatidylcholine (PC) differs from phosphatidylethanolamine (PE) by three methyl groups, a PC fragment resulting from their loss and substitution with H+ can coincide in m/z with the protonated PE molecule of the same acyl composition, so PC and PE cannot be told apart on positive-mode data alone. Negative mode resolves this, because PE produces the deprotonated molecule [M-H]- as its characteristic base peak, whereas PC does not and forms instead a set of characteristic fragment ions. As the [M-H]- is a necessary condition to confirm a PE, acquiring both polarities was required to distinguish the two classes. The added negative polarity therefore served to disambiguate the PL classes, not to cover a different lipid class: both polarities were applied to the same PL fractions, whereas the TAGs were analyzed in positive mode only, which was sufficient for their assignment.

What specific chromatographic advantages does ultra-high-performance liquid chromatography (UHPLC) offer over conventional HPLC when separating structurally similar intact lipid species like TAGs within a complex oil matrix?

MB: In our case, the main advantage of UHPLC mode was not a dramatic increase in separation compared to the original HPLC method since the original HPLC method already provided satisfactory TAG resolution. We transferred it from a 150 × 4.6 mm, 2.7 μm column to a 100 × 2.1 mm, 2.0 μm UHPLC column (both superficially porous particles) and adapted the gradient, flow rate, and injection conditions to maintain essentially the same chromatographic performance while significantly reducing analysis time and solvent consumption. An additional advantage of using 2 μm particles rather than sub-2 μm technologies is that similar chromatographic efficiency can often be achieved with lower backpressure, resulting in less stress on pumps, fittings, and column connections. This is particularly interesting because, for highly complex lipid matrices such as fish oils or dairy products, a single UHPLC column may still not provide enough peak capacity. If (HR)MS/MS systems such as orbital ion trap or quadrupole time-of-flight (QTOF) are available, they can compensate for residual co-elutions. However, when MS resolving power is limited, chromatography becomes even more important. In that context, 2 μm columns offer a practical advantage, providing near-UHPLC performance while leaving enough pressure headroom to connect multiple columns in series and increase peak capacity without excessively stressing the LC system. That gives a nice balance between efficiency, robustness, and flexibility.

After solid-phase extraction (SPE) fractionation isolated the free fatty acids (FFA) fraction, why was a separate derivatization-to-methyl-esters step necessary before gas chromatography (GC) analysis, and how did this affect the resolution or separation of individual fatty acid peaks?

MB: For the FFA fraction, we used a two-step derivatization in a single 2 mL vial. First, a base-catalyzed transesterification (KOH/MeOH) converted any residual esterified fatty acids (EFAs) into FAMEs. This step does not derivatize FFAs, so the FFAs remained untouched. This ensured that the resulting chromatogram truly represented the FFA pool without interference from residual lipid esters. We then applied an acid-catalyzed esterification (H₂SO₄/MeOH), which selectively converted only the FFAs into FAMEs for GC-FID analysis. Indeed, converting FFAs into alkyl derivatives improves chromatographic behavior, resulting in sharper peaks and better separation of individual FAs by GC. The same derivatization workflow can be applied directly to a total lipid extract, allowing separate characterization of EFAs and FFAs from the same sample.

When lipid classes with markedly different polarities—like neutral TAGs versus polar phospholipids—are run through the same non-fractionated extract, what specific coelution problems arise on the chromatographic column, and how does SPE cleanup eliminate them?

MB: The answer depends strongly on the chromatographic mode. For example, in HILIC mode, which provides a class-based separation, PLs are separated according to their polar head groups, so each class and subclass elutes as a distinct band, whereas neutral lipids such as TAGs elute together at the beginning of the run. In this mode, class overlap essentially does not occur.

In reversed-phase chromatography, which is the mode used in our work, retention follows the equivalent carbon number (ECN = carbon number (CN) − (2 × double bond number [DBN])), where the factor of 2 represents the cumulative contribution of all double bonds present in the molecule. Within a given lipid class, compounds with the same ECN can still be separated because retention is influenced not only by the total number of carbons and double bonds, but also by their distribution among the acyl chains, as well as by the position and geometry of the double bonds. Across different classes, however, the ECN ranges can overlap. For example, a PL with a high ECN may elute in the same retention region as a TAG with a low ECN. This can become relevant in highly complex matrices, such as milk-derived products or fish and marine organisms, which contain short-chain saturated and/or long-chain polyunsaturated fatty acids.

In pistachio, this overlap would not actually occur, because its FA profile does not include those specific FAs. Nonetheless, we separated the classes by SPE for two distinct reasons. The first was to optimize the chromatographic separation for each lipid class: TAGs were analyzed by non-aqueous reversed phase, while the PLs required dedicated mobile phases with a buffer, and running both classes in a single analysis would have compromised one or the other. The second reason went beyond chromatography. In fact, SPE was also used to isolate classes of specific interest. In our case, our goal was to investigate the possible presence of acidic PLs. So, SPE was not only a way to avoid a possible cross-class co-elution, but also to target minor classes.

How did you correlate retention time behavior on the UHPLC system with lipid class identity for TAGs and PLs, and did the quadrupole mass spectrometer (qMS) detector provide additional confirmation beyond retention time alone for resolving closely eluting lipid species?

GG: In reversed-phase mode, lipid classes eluted according to their ECN, so the elution order already provided a first useful indication. This relationship is class-dependent: at equal ECN, a PL, being more polar, elutes earlier than a TAG. Among PLs, retention further depends on the head-group polarity and molecular structure. Beyond retention, the qMS provided structural confirmation through the fragmentation generated in the APCI source. APCI generates the protonated/deprotonated molecular species together with characteristic in-source fragment ions arising from the loss of fatty acyl moieties and, for PLs, of the polar group, giving the corresponding diacylglycerol ion. For closely eluting species, whenever multiple structural possibilities were compatible with the same MS and chromatographic information, all plausible assignments were reported rather than selecting a single structure. The proposed acyl-chain compositions are therefore putative assignments supported by the combined interpretation of GC and MS data. Although APCI-qMS cannot provide the same level of structural characterization as dedicated ESI-MS/MS platforms, its broad availability, lower cost, and operational simplicity make it an attractive option for routine lipid characterization.

References

  1. Bellinghieri, C.; Giacoppo, G.; Schincaglia, A. et al. Applicability of MTBE-based Lipid Extraction Methods Assisted by Microwave in Food Analysis. Statistical Comparison and Greenness Evaluation with Soxhlet and Matyash. Talanta 2026, 299,129124. DOI: 10.1016/j.talanta.2025.129124
  2. Giacoppo, G.; Bellinghieri, C.; Murian, R. et al. Lipidomics Investigation of Sicilian Pistachio (Pistacia vera L. var. Bronte) by Microwave-assisted Extraction, Solid-phase Extraction/Fractionation, GC-FID, and UHPLC-APCI-MS Analyses. J Chromatogr A 2026, 1787, 467438. DOI: 10.1016/j.chroma.2026.467438