GC and LC–HRMS Track Laboratory Plastic Additive Leaching
Key Takeaways
- Syringe filters represented the largest NPPA source, with nylon filters releasing approximately threefold more additives than PTFE, underscoring filter polymer choice as a critical control point.
- Prepackaged QuEChERS salt kits introduced substantially higher additive backgrounds versus hand-weighed salts; HCPK contamination was roughly 1800-fold higher, risking major positive bias in trace assays.
Gas chromatography– (GC) and liquid chromatography–high-resolution mass spectrometry (LC–HRMS) reveal which laboratory consumables leach most plastic additives.
Plastic additives are chemicals mixed into plastics to make them softer, sturdier, or longer-lasting. Scientists are paying more attention to them because they linger in the environment and can harm health, so laboratories need to measure even tiny amounts of them accurately. There is a catch, however: the laboratory equipment used to prepare samples is often made of plastic, too, and it can leak these same chemicals into the samples. So far, little is known about how much contamination comes from everyday laboratory supplies and procedures.
To fill that gap, researchers at the University of Almería in Spain carried out what they believe is the first thorough study measuring how much plastic-additive contamination comes from laboratory supplies and chemicals. They focused on 28 additives other than phthalates, the best-known group of plastic additives. They ran “blank” tests, with no actual sample, through centrifuge tubes, pipette tips, syringes, syringe filters, and the salts used to pull chemicals out of samples, using two common lab solvents, ethyl acetate and acetonitrile. They then checked for contamination with gas chromatography and liquid chromatography, each paired with high-resolution mass spectrometry (GC-HRMS and LC-HRMS, respectively). The team published the work in the journal Analytica Chimica Acta.¹
Why Are Plastic Additives in Laboratory Supplies a Problem?
Plastic is everywhere in today’s laboratories; being flexible, inexpensive, and less likely to shatter (if at all) makes it essential for much of the everyday work. Glass can sometimes take its place, but not always, especially when a laboratory handles large numbers of samples or needs to prepare them in many steps. Most laboratory supplies are made from common plastics such as polypropylene, PET (the plastic in water bottles), and polycarbonate. To do their job, these plastics rely on added chemicals that help them stand up to heat, sunlight, wear and tear, and contact with other chemicals.2
Phthalates are a family of chemicals added to plastics, mainly to make them soft and flexible. Some have been linked to health problems, including effects on hormones. As rules limiting their use have grown stricter, manufacturers have increasingly switched to other additives instead. These replacements, non-phthalate plastic additives (NPPAs), are a varied group drawn from several chemical families, including citrates, adipates, and phosphate esters.3,4 Growing evidence shows that these replacement additives can also leak out of laboratory equipment and into samples, which can throw off measurements of very small amounts.5,6 Even so, scientists still know little about how much contamination they cause during everyday laboratory work.7
Which Laboratory Supplies Cause the Most Contamination?
The researchers report that syringe filters turned out to be the biggest source of contamination. Filters made of nylon released about three times as many plastic additives as filters made of Teflon (PTFE). Ready-made QuEChERS salt kits were much dirtier than salts weighed out by hand. One chemical, 1-hydroxycyclohexyl phenyl ketone (HCPK), a compound used to help plastics harden, showed up at levels roughly 1800 times higher in the kits. For most of the supplies tested, samples prepared with ethyl acetate picked up more contamination than those prepared with acetonitrile. Storage mattered too: a plastic-stored technical-grade acetone carried about 13 times more background contamination than a high-purity acetone kept in glass.
How Much Can Careful Choices Reduce Contamination?
When the researchers added up the contamination across a whole procedure, the choice of supplies made a huge difference. In gas chromatography workflows, the cleanest setup had about one-tenth the contamination of the dirtiest one. In liquid chromatography workflows using QuEChERS, the gap was more than a hundredfold. In other words, choosing laboratory supplies carefully can greatly cut the amount of contamination that builds up during sample preparation.
“This study,” write the authors of the paper,1 “provides the first quantitative assessment of NPPA contamination arising from routine laboratory consumables across complete sample-preparation workflows. The results identify critical contamination sources, establish practical mitigation strategies, and demonstrate that consumable selection can have a greater impact on NPPA background levels than the extraction workflow itself, thereby improving the reliability of trace-level plastic additive determinations.”
Read More on Similar Topics
References
- Capilla-Flores, R.; López-Ruiz, R.; Romero-González, R. et al. Non-phthalate Plastic Additives Contamination from Laboratory Consumables: A Quantitative Assessment Across Sample Preparation Workflows. Anal Chim Acta 2026, 1428, 346452. DOI:
10.1016/j.aca.2026.346452 - Hermuth-Kleinschmidt, K. The Different Plastics Used in the Laboratory, White Paper no. 79, Hamburg, Germany. Sigma-Aldrich website.
https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/215/657/application-note-079.pdf (accessed 2026-10-09) - Wiesinger, H.; Wang, Z.; Hellweg, S. Deep Dive into Plastic Monomers, Additives, and Processing Aids. Environ. Sci. Technol.2021, 55, 9339-9351. DOI:
10.1021/acs.est.1c00976 - Flora, G.; Stephy, G. M.; Veeramuthu, A. The Rise of Non-phthalate Plasticizers: Serious Risks to Human Life and Environmental Consequences - A Comprehensive Review. J. Environ. Chem. Eng.2025, 13, 115976. DOI:
10.1016/j.jece.2025.115976 - Yang, J.; Li, Y.; Wang, Y. et al. Recent Advances in Analysis of Phthalate Esters in Foods. Trends Anal. Chem.2015, 72, 10-26. DOI:
10.1016/j.trac.2015.03.018 - Russo, M. V.; Avino, P.; Perugini, L. et al. Extraction and GC-MS Analysis of Phthalate Esters in Food Matrices: A Review. RSC Adv.2015, 5, 37023-37043. DOI:
10.1039/c5ra01916h - Fuente-Ballesteros, A.; Obaydo, R. H.; Elagamy, S. H. et al. Cross-contamination Pathways in the Analysis of Plastics and Related Chemical Compounds: Good Laboratory Practices and Tips. Trends Environ. Anal. Chem.2026, 50, e00304. DOI:
10.1016/j.teac.2026.e00304
Related to this article