HPLC 2017 Topic Preview: The Role of LC–MS in Lipidomics

This is the sixth instalment of a series of articles exploring current topics in separation science that will be addressed at the HPLC 2017 conference in Prague, Czech Republic, from 18–22 June.

This is the sixth instalment of a series of articles exploring current topics in separation science that will be addressed at the HPLC 2017 conference in Prague, Czech Republic, from 18–22 June.

Lipidomics, the analysis of lipids by mass spectrometric methods, revolutionized lipid science (1). It provides detailed quantitative information on hundreds of lipid species and opens new possibilities to gain an insight into lipid biology. This helps not only to explain the vital role of lipid species as membrane building blocks, but also to unravel their bioactive functions. Thus, lipid species can act as signalling molecules and modulate membrane properties, which are essential for organelle and membrane protein function. Moreover, the first examples demonstrated their potential as novel biomarkers to monitor human health (2).    Lipidomics research is based on two main approaches: direct infusion mass spectrometry (MS) analysis (shotgun lipidomics) and liquid chromatography (LC)–MS analysis. In direct infusion analysis a crude lipid extract is infused into the mass spectrometer and lipid species identification relies on specific precursor ion, neutral loss scans (3). The main advantage of infusion-based analysis is its simplicity and the straight-forward way of quantification. Analytes and internal standards are present in the same sample matrix and thus experience the same ion suppression and matrix effects. Shotgun analysis is therefore able to provide comprehensive, quantitative lipidomes, as for example demonstrated for yeast (4). The application of high mass resolution, MSn, and derivatization–gas phase reactions can provide detailed lipid structures. However, the application of shotgun approaches is limited in sensitivity and separation of isomeric lipid species. Moreover, in-depth characterization of lipid species in crude lipid extracts may be complicated by co-isolation of precursor ions.    LC–MS provides separation of lipids and reduces the complexity of the matrix. It typically provides a higher sensitivity than shotgun and offers retention time as an additional parameter to identify lipid species. Therefore, low abundant lipid mediator species are typically analyzed by targeted LC–tandem mass spectrometry (MS/MS) (5). Lipidomic methods apply both reversed phase with nonpolar as well as normal phase and hydrophilic interaction chromatography (HILIC) with polar selectivity.    Reversed phase separates lipids species based on their hydrophobic moiety, that is, the hydrocarbon chain for most lipid classes. This allows the separation of hundreds of lipid species based on their chain lengths and double bond number (6,7). The retention behaviour follows certain rules and increases the confidence of lipid species identification. This permits the separation of isomeric lipid species with different acyl chains like PC 18:1_18:1 (phosphatidylcholine with two acyl chains containing 18 carbon atoms and 1 double bond) and PC 18:0_18:2 (18:0 and 18:2 acyl chain). However, quantification in lipidomics typically relies on lipid species not present in the samples. In reversed-phase chromatography most of the lipid species and internal standards elute at different times, thus experience different matrix effects and different solvent composition, which influences their ionization and may result in inaccurate quantification (8).    The polar selectivity of normal phase and HILIC provides lipid class-specific separation. This has great advantages in terms of quantification since analytes and internal standards show similar retention times. Moreover, identification of lipid species comprising a lipid class is straightforward. In contrast to reversed-phase methods, separation of acyl chain isomers is not usually possible by normal phase and HILIC methods, but lipid class-specific separation may resolve isomers like bis(monoacylglycero)phosphate and phosphatidylglycerol (9). A promising approach is the application of polar stationary phases with ultrahigh-performance supercritical fluid chromatography (UHPSFC), which  offers ultrafast separation for quantitative analysis of multiple lipid classes (10).   Today, an increasing number of studies are reporting poor quality lipidomics data with misidentification and inaccurate or inappropriate quantification of lipid molecules. These studies primarily use untargeted metabolomics approaches (11) and the reasons for the poor data quality include analytical, bioinformatics, and educational aspects. Therefore, it is necessary to implement reporting standards for lipidomics data to share with the scientific community (12). These standards need to cover both shotgun and LC–MS approaches. Only the application of both approaches in a complementary and confirmatory way permits a comprehensive and accurate coverage of the lipidome.  

References

  (1) K. Yang and X. Han,

Trends Biochem. Sci.

41

(11), 954–969 (2016). (2) S. Sales

et al

.,

Sci. Rep.

6

, 27710 (2016). (3) X. Han, K. Yang, and R.W. Gross,

Mass Spectrom. Rev.

31

(1), 134–78 (2012). (4) C.S. Ejsing

et al

.,

Proc. Natl. Acad. Sci. U.S.A

106

(7), 2136–2141 (2009). (5) G. Astarita

et al

.,

Biochima et Biophys Acta

1851

(4), 456–68 (2015). (6) M. Ovcacikova

et al

.,

J. Chromatogr. A

1450

, 76–85 (2016). (7) K. Sandra and P. Sandra,

Curr. Opin. Chem. Biol.

17

(5), 847–53 (2013). (8) S. Krautbauer, C. Buechler, and G. Liebisch,

Analytical Chemistry

88

(22), 10957–10961 (2016). (9) M. Scherer, G. Schmitz, and G. Liebisch,

Analytical Chemistry

82

(21), 8794–8799 (2010). (10) M. Lisa and M. Holcapek,

Analytical Chemistry

 

87

(14), 7187–95 (2015). (11) G. Liebisch, C.S. Ejsing, and K. Ekroos,

Clinical Chemistry

61

(12), 1542–1544 (2015). (12) G. Liebisch

et al

.,

Biochimica et Biophysica Acta

(2017).


Related to this article

Best of the Week: From Cancer Diagnostics to Smarter LC Method Development
This week, Chromatography Online’s coverage spanned lipidomics-based early cancer detection, career paths in separation science, machine learning and uncertainty-driven approaches to LC method development, new EU wastewater rules for pharmaceutical micropollutants, automated Separation Quality Factor (SQF) evaluation in chromatography data systems, and solutions to GLP-1 peptide analysis challenges.