LCGC Advances In Gas Chromatography 11.23-11.24
|Articles|June 1, 2011

The Application Notebook

  • The Application Notebook-06-01-2011
  • Volume 0
  • Issue 0

An Improved SPE-LC–MS–MS Platform for the Simultaneous Quantification of Multiple Amyloid Peptides in Cerebrospinal Fluid for Preclinical or Biomarker Discovery

Fast, flexible platforms for peptide quantification are needed, particularly for a discovery setting. This type of methodology would be especially advantageous in the case of amyloid beta (a?) peptides.

Erin E. Chambers1, Mary E. Lame2, and Diane M. Diehl1, 1Waters Corporation and 2Pfizer, Neuroscience Research Unit

Fast, flexible platforms for peptide quantification are needed, particularly for a discovery setting. This type of methodology would be especially advantageous in the case of amyloid beta (aβ) peptides. The deposition/formation of insoluble aggregates, or plaques, of aβ peptides in the brain is considered to be a critical event in the progression of Alzheimer's disease (AD) and thus has the attention of many researchers. A previous Waters application note (720003682en) described in detail the development of a fast, flexible SPE-LC–MS–MS platform for the quantification of multiple aβ peptides from human or monkey CSF for use in a biomarker or preclinical discovery setting. In this work, the mass spectrometry platform has been updated from the Xevo TQ MS to the Xevo TQ-S mass spectrometry system. This change facilitated both a 4× reduction in required sample size and a 4–5× increase in assay sensitivity. This work focuses on methods for the 1–38, 1–40, and 1–42 aβ, Table I.

Table I: Sequence, MW, and pI Information for Amyloid Peptides

Experimental Conditions

SPE-LC–MS–MS Conditions

LC system: Waters ACQUITY UPLC System

Column: ACQUITY UPLC BEH C18 300 Å, 2.1 × 150 mm, 1.7 µm, Peptide Separation Technology

SPE: Oasis MCX µElution 96-well plate, 50 µL human or animal CSF

MS system: Waters Xevo TQ-S, ESI+

Figure 1: Representative ESI+ MS-MS spectrum for amyloid 1-42 with fragment sequence ions labeled.

Results and Conclusions

  • An improved SPE-UPLC–MS–MS bioanalytical method was developed and validated for the simultaneous quantification of multiple amyloid β peptides in human CSF.

  • MS was performed in positive ion mode since CID of the 4+ precursor ion yielded several distinct product ions corresponding to inherently specific b sequence ions (representative spectrum shown in Figure 1).

  • UPLC separation of the three amyloid β peptides is shown in Figure 2.

Figure 2: Representative UPLC–MS–MS analysis of amyloid 1-38, 1-40, and 1-42 peptides extracted from artificial CSF + 5% rat plasma.

  • The increased sensitivity of the Xevo TQ-S triple quadrupole mass spectrometer facilitated the use of 4× less sample and a 4–5× improvement in quantification limits (Table II).

  • Average basal levels and RSD values for all 3 aβ peptides in two sources of human CSF are shown in Table III and are lower or equal to 5%.

  • Overspiked QC samples were prepared in triplicate in two sources of pooled human CSF at 0.04, 0.075, 0.15, 0.2, 0.8, 2, and 6 ng/mL. Accuracy and precision values met the regulatory criteria for LC–MS–MS assays. Results from QC sample analysis are shown in Table IV. Average deviation from expected is 2.3%.

  • The method described herein eliminates time-consuming immunoassays or immunoprecipitation steps for preclinical work.

  • The use of a single UPLC–MS–MS assay represents a significant advantage over an ELISA assay, which would require multiple assays with multiple antibodies to quantify each of the relevant peptides.

Table II: Comparison of Standard Curve and QC Range Using Xevo TQ and TQ-S MS

Table IV: Average Deviation Values for all Overspike QC Samples

Table III: Baseline Levels of Amyloid β Peptides in Two Sources of Pooled Human CSF

Copyright 2010: ACQUITY UPLC, Oasis, The Science of What's Possible, Xevo are trademark of Waters Corporation.

Waters Corporation

34 Maple Street, Milford, MA 01757

tel. (508) 478-2000, fax (508) 478-1990

Website: www.waters.com


Related to this article