LCGC Advances In Gas Chromatography 11.23-11.24
|Articles|December 2, 2013

The Application Notebook

  • The Application Notebook-12-02-2013
  • Volume 0
  • Issue 0

Soluble Fraction Analysis in Polypropylene for Quality Control Laboratories

Polymer Char

Introduction

The analysis of the amorphous fraction is a fundamental task in the production of polypropylene (PP). It provides a measurement of the small amounts of undesirable low tacticity and low molar mass fractions in homopolymer resins, which have significant influence on the polymer properties and processing. The incorporation of amorphous material is usually done within the same production plant at a second stage of the polymerization process by adding ethylene to the continuous reaction, resulting in the formation of amorphous ethylene-propylene copolymer. The continuous measurement of the overall amount of this rubbery fraction is essential for plant control in order to obtain the desired specifications and performance of the polymer resin.

Historically, the amorphous or soluble content has been measured by solubility in xylene using manual wet chemistry methods that are time-consuming and involve large amounts of solvent. CRYSTEX QC is an innovative approach that automates the analysis of PP samples in quality control (QC) plant or processing laboratories. In addition, CRYSTEX QC quantifies ethylene incorporation and intrinsic viscosity in the original sample and in both the amorphous and crystalline fractions.

Classic Gravimetric Method

This traditional method is based on a crystallization process in which the crystalline matrix is segregated from the amorphous or soluble fraction by cooling a xylene solution according to a specified method with fixed crystallization temperature and time (1–2). The process is described as follows:

"A weighed amount of polymer is dissolved in hot xylene, then cooled under controlled conditions down to 25 °C, which results in the precipitation of the insoluble fraction. The soluble matter remains in the xylene. The suspension is filtered and an aliquot of the solution is then evaporated, dried, and the residue weighed. The ratio of the corresponding weights is the basis for the final calculation of the soluble fraction."

This procedure has been extensively used in the industry and has proven to be of great value to the manufacturing of PP. However, the determination of the xylene solubles with this approach demands significant manpower, solvent consumption, and an analysis time of around 5–6 h. The ethylene content of the soluble fraction is measured separately by FTIR on a pressed film; and measurement of intrinsic viscosity demands re-dissolution of the dried fraction and injection into a viscometer.

New Automated Crystallization Method

CRYSTEX QC is a fully automated instrument that quantifies the amorphous fraction while delivering improvements in time, safety and precision. The instrument is based on a proprietary Temperature Rising Elution Fractionation (TREF) column where a small aliquot of homogeneous polymer solution is crystallized, in di- or tri-chlorobenzene, on a support under reproducible and well-controlled conditions, therefore, delivering very reproducible results. Solvent selection is not so important as the intention to separate two different polymer phases quite apart in crystallinity (3) and it has been proven by previous CRYSTEX methodology (4–5)

Figure 1: CRYSTEX QC for process and QC laboratories of polypropylene manufacturing plants.

The amorphous fraction is measured with a sensitive infrared (IR) detector that delivers equivalent values to the xylene solubles measurement but obtained with outstanding precision. The IR detector also measures the concentration of the amorphous and crystalline fractions, as well as their ethylene incorporation. The integration of a capillary viscometer provides an automated measurement of intrinsic viscosity of the whole sample and of both the amorphous and crystalline fractions.

Figure 2: Schematic diagram of CRYSTEX QC.

The schematics of the system are presented in Figure 2. All the analyst is required to do is put an approximate amount of sample in a disposable bottle, place it in the stirred-heated plate, and lower a handle to pierce the septum of the bottle with a needle. The automated process is as follows:

a) Filling the disposable bottle with 100 mL of solvent preheated at 160 °C. Initiate stirring in the heated plate.

b) An aliquot of the solution is pumped through the TREF column at 160 °C into the detectors to measure concentration, content of ethylene, and intrinsic viscosity of the whole sample (first eluted peak in Figure 3).

c) A new aliquot of the solution is injected into the middle of the column at 160 °C, the flow is stopped, and the column temperature is reduced down to controlled ambient temperature to precipitate the crystalline fraction. The column is then flushed to elute the soluble fraction (second peak in Figure 3). The ratio of this peak area by the one of the whole sample provides the soluble percentage in the sample. As the soluble material passes through the IR and viscometer detectors, ethylene content and intrinsic viscosity of this fraction are also measured.

d) Temperature is raised rapidly up to 160 °C with stop flow to re-dissolve the crystalline fraction and after a pre-set time, this fraction is eluted through both IR and viscometer detectors.

e) The column is rinsed, and the instrument is ready for a new sample analysis.

Figure 3: Elution of the whole sample and PP fractions in a TREF column.

Results and Discussion

Reproducibility is remarkable for the soluble fraction determination. Table 1 shows data obtained from seven replicate analyses using 4 g of sample for a set of three PP products. Average standard deviations are shown for each type of measurement.

Table 1: Soluble fraction, ethylene content, and intrinsic viscosity of three different PP products.

No additional experimental effort is required since the data are collected by the IR and viscometer detectors during the automated analysis and a measurement of the ethylene content and intrinsic viscosity of all polymer parts is provided.

Conclusions

A new method that measures the soluble fraction in PP resins has been developed based on a TREF crystallization approach. The analytical process is fully-automated with no need for weighing, filtration, or solvent handling.

The total time of analysis is 2 h, and samples are analyzed one at a time in disposable bottles. Total volume consumption is as low as 180 mL for a 2 g sample analysis. Analysis of 4 g of resin in the form of powder can be done for enhanced sample representativeness.

In addition to the soluble fraction, the new system measures ethylene content and intrinsic viscosity simultaneously in both amorphous and crystalline fractions, and in the whole sample.

References

(1) ASTM standard D5492-10, ASTM.

(2) ISO standard 6427, (1992), ISO standard 16152, (2005).

(3) B. Monrabal, Encyclopedia of Analytical Chemistry (John Wiley & Sons, 2000) pp. 8074–8094.

(4) L. Romero, B. Monrabal, and A. Ortín, Pittcon 2001 USA.

(5) A. Ortín, B. Monrabal, and M. D. Romero, LCGC Europe Supplement 19(3),, 32 (2006).

Polymer Characterization S.A.

Gustave Eiffel 8, Paterna, Valencia E-46980 Spain

Tel: +34 96 131 81 20 fax +34 96 131 81 22

E-mail: info@polymerchar.com

Website: www.polymerchar.com


Related to this article

Magnetic Aerogel Extracts Pesticides for LC–MS/MS Testing
Kazhal Khazaei of K.N. Toosi University of Technology (Iran) discusses her research team’s development of a magnetic copper-based metal-organic gel aerogel that captures tiny amounts of pesticides from plant-based milks for liquid chromatography.–tandem mass spectrometry (LC–MS/MS) testing with LCGC International.