Analysis of L-Theanine in Tea by HPLC with Post- Column Derivatization

Homogenize 2 g of dry tea leaves with 25 mL of Li220 for 5 min. Centrifuge and filter through 0.45 µm nylon filter.

Theanine is a neurologically active amino acid found in tea plants. L-Theanine is a dominant amino acid in green tea and is responsible for its unique pleasant taste as well as known relaxation effect. Theanine also has been proven to reduce physical and mental stress, and improve cognition and mood.


Table I: Column oven program

Cation-exchange chromatography using post-column Ninhydrin reagent Trione and UV detection has shown unmatched reproducibility and selectivity in the analysis of free amino acids in complex matrices. The Pinnacle PCX post-column derivatization system allows shortening run times by utilizing column temperature gradients.


Table II: HPLC gradient

We introduce a simple and robust method for analysis of L-Theanine in addition to the other free amino acids in tea leaves.


Figure 1: Chromatogram of green tea sample.

Method

Sample Preparation

Homogenize 2 g of dry tea leaves with 25 mL of Li220 for 5 min. Centrifuge and filter through 0.45 µm nylon filter.


Click here to view full-size graphic


Figure 2: Chromatogram of white tea sample.

Pickering Laboratories, Inc.
1280 Space Park Way, Mountain View, CA 94043
tel. (800) 654-3330, (650) 694-6700
Website: www.pickeringlabs.com

Articles in this issue

almost 11 years ago

ACCTA, Inc.

almost 11 years ago

Benson Polymeric, Inc.

almost 11 years ago

CryoXtract Instruments, LLC

almost 11 years ago

Agilent Technologies Inc.

almost 11 years ago

Ellutia

almost 11 years ago

Chem Service, Inc.

almost 11 years ago

FMS, Inc.

almost 11 years ago

Hamilton Company

almost 11 years ago

GERSTEL, Inc.

Related to this article