Analysis of Limonin in Citrus Juice Using QuEChERS and LC–MS/MS
This application note outlines a simple, fast, and cost-effective QuEChERS-based method for the determination of limonin in citrus juice. Limonin is extracted from a variety of juice samples using acetonitrile and citrate-buffered salts. The sample extract undergoes cleanup by dispersive-SPE (dSPE) using primary-secondary amine (PSA), C18, and graphitized carbon black (GCB) to remove unwanted matrix components, including sugars, acids, and pigments, and yield a clear sample extract. Analysis is performed by LC–MS/MS using a Selectra® C18 HPLC column (although HPLC-UV can also be used).
Procedure
1. Sample Extraction:
a) Allow citrus juice to come to room temperature and thoroughly mix the sample before analysis.
b) Transfer 10 mL of juice into a 50 mL polypropylene centrifuge tube.
c) Add 10 mL of acetonitrile.
d) Add the contents of the ECQUEU7-MP Mylar pouch and vortex or shake (manually or mechanically) for 5 min.
- For this work a SPEX® SamplePrep® 2010 Geno/Grinder® was used (1500 rpm).
e) Centrifuge the samples at ≥3000 rcf for 5 min.
2. Sample Clean-up of Extract
a) Transfer 1 ml of supernatant to a CUMPSC1875CB2CT 2 mL dSPE tube.
b) Vortex the samples for 30 s.
c) Centrifuge the samples at ≥3000 rcf for 5 min.
d) Transfer 500-600 µL of purified supernatant into an autosampler vial for analysis.
Instrumental
Results
Samples were fortified with limonin at 10 µg/mL (50 µL of a 2 mg/mL acetonitrile standard). A six-point matrix-matched calibration curves were prepared at 1, 2, 5, 10, 15, and 20 µg/mL. Non-fortified juice samples were extracted according to the QuEChERS procedure described above and 1 mL of purified extracts were fortified with 5, 10, 25, 50, 75, and 100 µL of a 200 µg/mL acetonitrile standard.
* One outlier was omitted from the final results.
Conclusion
This application note outlines a simple, fast, and cost-effective QuEChERS-based method for the determination of limonin in citrus juice. Citrus juice samples are extracted with citrate-buffered QuEChERS salts followed by dSPE cleanup of the supernatant using PSA, C18, and GCB, yielding a clear extract. Analysis is performed by LC–MS/MS using a Selectra® C18 HPLC column. Various citrus juices (lemon, lime, orange, and grapefruit) were evaluated for recovery and repeatability using fortified samples. Excellent recovery (88 to 97%) and repeatability (≤ 7.5%) were obtained using the outlined method.
UCT, LLC
2731 Bartram Rd, Bristol, PA 19007
tel: (800) 385-3153, Email:
Website:
Articles in this issue
over 9 years ago
Ensuring Protein Regent Quality by SEC-MALSalmost 10 years ago
Vol 35 No 2 LCGC North America February 2017 Application Notebook PDFalmost 10 years ago
A New Silica-Based Aqueous SEC Column for Analysis of Antibody Proteinsalmost 10 years ago
Rapid Analysis of C3 Epimers of 25-Hydroxyvitamin D in Serum by LC–MS/MSalmost 10 years ago
Characterization of PLGA Using SEC-MALS-IValmost 10 years ago
Arsenic Speciation on Hamilton PRP-X100Related to this article