Ask the Editor: GPC-SEC

Article

E-Separation Solutions

E-Separation SolutionsE-Separation Solutions-09-10-2008
Volume 0
Issue 0

The following question and answer appeared in an article by Daniela Held in the June 2008 issue of The Column, LCGC Europe’s e-zine. Why does GPC–SEC require a calibration?

The following question and answer appeared in an article by Daniela Held in the June 2008 issue of The Column, LCGC Europe’s e-zine.

Why does GPC–SEC require a calibration?

The primary information obtained from gel permeation chromatography (GPC)–size-exclusion chromatography (SEC) standard detectors (for example, UV, evaporative light scattering, or refractive index) is not the molar mass but the apparent concentration at a certain elution volume. It is only with the GPC–SEC calibration curve and the concentration profile from the concentration detector that all molar mass averages and the molar mass distribution can be obtained (1,2). GPC–SEC is, therefore, a relative method. The GPC–SEC calibration is based on assigning the elution volume to a molar mass. This is in contrast to high performance liquid chromatography (HPLC), where the detector response (signal intensity, peak area) is calibrated and assigned to a concentration.

Another possibility to measure the relation of the molar mass to elution volume is to apply an on-line light scattering detector (low-angle, right-angle, and multiangle laser light scattering but not evaporative light scattering) and a concentration detector (UV, refractive index, evaporative light scattering). With such a combination a calibration curve valid for a specific sample is measured with the sample itself. To obtain smooth molar mass distributions, light-scattering software programs often provide the use of fitted data, where (same as in conventional calibration curves) the molar mass/elution volume relation is fitted. Proper data analysis also requires the detector constant (calibration constant) of the light-scattering detector and of the concentration factor (response factor). For the concentration detector, the constant is comparable to detector calibrations performed in HPLC and obtained in a similar way.

(1) D. Held, The Column, 8 (August 2007).

(2) D. Held and G. Reinhold, The Column, 10 (October 2007).

Questions?

LCGC technical editor Steve Brown will answer your technical questions. Each month, one question will be selected to appear in this space, so we welcome your submissions. Please send all questions to the attention of "Ask the Editor" at lcgcedit@lcgcmag.com We look forward to hearing from you.

Related Videos
Toby Astill | Image Credit: © Thermo Fisher Scientific
Robert Kennedy