Blood Alcohol Content Analysis Using Nitrogen Carrier Gas

Article

The Application Notebook

The Application NotebookThe Application Notebook-09-01-2015
Issue 0

 

Alcohol consumption can seriously affect the ability of a driver to operate a vehicle and blood alcohol content (BAC) directly correlates with this impairment. A number of nations have zero alcohol tolerance for motorists, but the majority of countries worldwide have a limit of between 50 and 80 mg alcohol per 100 mL blood, or 0.05–0.08%. Results are used in court to provide quantitive levels of BAC, which makes it one of the most commonly practised analyses in forensic laboratories. The large number of samples and requirement for speed of sample processing means that analysis needs to be conducted quickly, whilst also giving reliable and accurate results. 

For analysis of BAC, headspace gas chromatography (GC) with flame ionization detection (FID) is typically used. Headspace GC allows the quantitative analysis of alcohol directly from blood samples. Standard headspace systems use nitrogen for vial presurization, with helium typically used for GC carrier gas. This application note looks at the use of nitrogen for both vial pressurization and GC carrier gas. Nitrogen offers a cost-effective, abundant alternative to helium for carrier gas, whilst also providing a similar performance. Here we compare analysis of real forensic blood samples, taken from motorists suspected of driving under the influence of alcohol, analyzed using nitrogen and helium carrier gas.

Sample Preparation
Using a Hamilton Microlab 600 Diluter, 200 μL of calibrators, controls, or blood samples were aliquoted and dispensed with 2000 μL of internal standard solution into a 10 mL headspace vial and capped. The internal solution consisted of 0.03% (v/v) n-propanol/1M ammonium sulphate/0.1 M sodium hydrosulphite. NIST traceable aqueous ethanol solutions from Cerilliant and Lipomed were used as calibrators (10, 50, 80, 200, 300, 500 mg/dL) and controls (20, 80, 400 mg/dL), respectively.

Experimental
Analyses were conducted using an Agilent 7890B GC with split/splitless inlet and dual columns each connected to an FID detector. Splitting of the samples onto the columns was via an Agilent unpurged Capillary Flow Technology splitter. The GC was coupled with an Agilent 7697A headspace sampler. Vial pressurization gas for all tests was provided by a Peak Scientific Precision Nitrogen Generator. Carrier gas was provided by either the helium cylinder or the Precision Nitrogen Standard Generator. The HS–GC–FID system operating condtitions are displayed in Table 1.

 

 

The software used for analysis was Agilent MassHunter GC/MS Acquisition and MSD ChemStation Enhanced Data Analysis E.02.02.1431

Results
Calibration curves produced with helium and nitrogen carrier gas both gave very good linearity with both curves having R2 values of 99.9999 (Figure 1). Blood alcohol levels of five blood samples were analyzed.

 

Figure 1: Calibration curves for ethanol standards run using nitrogen and helium carrier gas.

 

Figure 2: Results of resolution mixture run on a DB-ALC1 column using nitrogen and helium carrier gas.

Figure 3: Results of resolution mixture run on a DB-ALC2 column using nitrogen and helium carrier gas.

 

Figures 2 and 3 show chromatograms from the DB-ALC1 and DB-ALC2 columns, respectively for the separation and elution order of analytes for the multi-component resolution mix when run using nitrogen and helium carrier gas. Separation of potentially interfering components such as methanol and 2-propanol was achieved within 3 min when using either carrier gas (Figure 2 and Figure 3).

 

Results of analyses of real blood samples (analyzed in duplicate) run with both nitrogen and helium carrier gas gave equivalent results with no differences found in the calculated ethanol concentrations (Table 2). Of the five blood samples tested, one was over 0.2%, which would result in a driving ban in almost every country worldwide. Two samples were over 0.05%, which would result in a driving ban in a number of countries. The other two samples were 0.014% and 0.023%, which would be below the limit in the majority of countries worldwide.

Conclusions
Results of BAC analysis show that there is no difference in the linearity of the calibration curve, or of the calculated ethanol content of real blood samples regardless of whether nitrogen or helium carrier gas is used.

As an abundant, inexpensive alternative to helium, which is becoming increasingly more costly, there is no reason why nitrogen cannot be used for BAC analysis in place of helium. Since nitrogen is often used for vial pressurization in headspace samplers, the use of a single gas source for vial pressurization, carrier gas and FID make-up gas simplifies the lab’s gas sourcing and would allow total gas supply from gas generators if the precision nitrogen was used in conjunction with the Precision hydrogen and zero air generators for GC–FID analysis.

 

Peak Scientific
Fountain Crescent, Inchinnan Business Park, Inchinnan, PA4 9RE, Scotland, UK
E-mail: info@peakscientific.com
Website: www.peakscientific.com


 

Related Videos
Robert Kennedy
John McLean | Image Credit: © Aaron Acevedo
Related Content