Rapid Determination of DNPH Derivatized Carbonyl Compounds by UHPLC

Article

The Application Notebook

The Application NotebookThe Application Notebook-07-02-2010
Volume 0
Issue 0
Pages: 8

Carbonyl compounds, in particular aldehydes, are reactive volatile substances. They are of concern to the public, because they are emitted as air pollutants by a large range of industrial processes and other combustion sources.

Silvia Marten and Mareike Naguschewski, KNAUER, Berlin, Germany.

Introduction

Carbonyl compounds, in particular aldehydes, are reactive volatile substances. They are of concern to the public, because they are emitted as air pollutants by a large range of industrial processes and other combustion sources. In addition, they can also be found indoors when emitted from sources such as insulation, furniture or tobacco smoke.1,2 Because of their adverse health effects, monitoring of these substances is important. Formaldehyde and acetaldehyde are known for their irritating effects on animals and humans whereby formaldehyde is also carcinogenic.3

Low molecular weight aldehydes are highly volatile and polar substances resulting in a hindered analysis by RP-UHPLC methods. In addition, many known carbonyl compounds have no chromophores and are for this reason not detected by UV. Due to their functional carbonyl group, low molecular weight aldehydes and ketones can, however, be derivatized with 2,4 dinitrophenylhydrazine (DNPH). In acidic media, DNPH reacts with the carbonyl group to form stable hydrazones (DNPH-carbonyls) with a lower vapour pressure that can be separated by UHPLC on a C18 column and detected easily by UV.2

In this work, six DNPH derivatized low molecular weight carbonyls are separated by UHPLC in less than 3 minutes. Optimizing the speed and resolution of routine analyses by applying a UHPLC method can not only save time but also dramatically decrease eluent costs, particularly important for analyses using acetonitrile.

Figure 1:

Experimental

A very fast method for identifying six DNPH-derivatized carbonyls in a mixture was possible using the Knauer PLATINblue UHPLC system. Employing the BlueOrchid C18 A phase with a 1.8 µm particle size allowed for the reduction of analysis time from about 20 to less than 3 minutes compared to a conventional HPLC method. A binary high pressure gradient configuration at a flow rate of 0.8 mL/min in combination with a 2 mm column i.d. and UV detection was applied. After measuring a sample containing 2 ng/µL DNPH derivatized carbonyls, calibration in the range of 2 up to 40 ng/µL was performed.

Method parameters

Column........ BlueOrchid C18 A 1.8 µm, 100 × 2 mm

Eluent A....... Water

Eluent B........Acetonitrile

Gradient

Time (min).... %A...... %B

0.0................60....... 40

2.0................45....... 55

3.0................0....... 100

3.5................0....... 100

Flow rate.......... 0.8 mL/min

Injection volume....... 2 µL standard

Column temperature....... 40 °C

Detection....... UV at 370 nm (50 mm cell, 50 Hz, 0.01 s)

Results and Conclusion

By means of derivatization with DNPH, six carbonyl compounds could be easily determined in less than 3 minutes by employing a Knauer PLATINblue UHPLC system, a BlueOrchid C18 A stationary phase and an acetonitrile elution gradient. The 2 mm inner diameter of the chosen column resulted in a comparable small amount of required eluent. All carbonyls are baseline separated with resolution values in the range of 2.5 up to 7.3. The analysis time could be reduced more than six times compared to the method using a conventional HPLC system. The limits of detection (LOD) lie in the range of 0.1 ng for all six carbonyls. Calibration is realized for all analysed compounds and the linearity (r2 ) was in the range of 0.999131–0.999916. By using UHPLC and its advantages, long equilibration and analysis times can be avoided, and a UV detection of DNPH-derivatized carbonyl concentrations in the range of 4 ng can be realized.

References

1. M. Possanzin et al., Chromatographia, 23(11), 829–834 (1987).

2. Y. Feng and J. Zhu, Anal. Sci., 20, 1691–1695 (2004).

3. J. Zhang, Environ. Sci., Technol., 28(1), 146–152 (1994).

Knauer GmbH

Hegauer Weg 38, 14163 Berlin, Germany

tel. +49 30 809727 0 fax +49 30 801501 0

E-mail: info@knauer.net Website: www.knauer.net