Thyroid Hormones in Serum and Plasma Using SPE Extraction and UHPLC–MS-MS Analysis
This application note describes the development of an analytical method for the sensitive and accurate determination of free and total thyroid hormones in serum and plasma using a mixed-mode SPE extraction procedure and UHPLC–MS-MS analysis.
This application note describes the development of an analytical method for the sensitive and accurate determination of free and total thyroid hormones in serum and plasma using a mixed-mode SPE extraction procedure and UHPLC–MS-MS analysis. Chromatographic separation, including baseline resolution of triiodothyronine and reverse triiodothyronine, is conducted in 3 min using a Selectra® DA column. Ultra-low hormones and steroids serum (Golden West Biologicals, Inc.®) were used to generate accuracy and precision data. The method was evaluated at two concentrations (0.2 and 1 ng/mL) using procedures for free and total hormones. Recoveries of the three hormones ranged from 87% to 100% and reproducibility was ≤3.5%.
Extraction/Analytical Materials
Procedure
1. Sample Preparation for free hormones:
(a) To 1 mL of plasma add 1 mL of 0.1 M HCl.
(b) Vortex for 30 s.
Sample Preparation for total hormones
(a) To 1 mL of plasma add 2 mL of acetonitrile.
(b) Vortex briefly and let stand for 5 min.
(c) Centrifuge at ≥1500 rcf for 5 min.
(d) Transfer supernatant to a clean glass tube.
(e) Add 2 mL of 0.1 M HCl and vortex for 30 s
2. Condition Cartridge
(a) 1 × 3 mL methanol.
(b) 1 × 3 mL 0.1 M HCl.
(c) Apply the sample to the SPE cartridge.
3. Wash Cartridge
(a) 1 × 3 mL 0.1 M HCl.
(b) 1 × 3 mL methanol.
(c) Dry cartridges for ~30 s to remove excess methanol.
4. Elution
(a) Elute with 1 × 3 mL methanol containing 2% ammonium hydroxide.
(b) Evaporate to dryness/Reconstitute in 100 µL of methanol.
Instrumental
LC–MS-MS: Thermo Scientific Dionex Ultimate 3000® LC System/ Thermo Scientific TSQ Vantage tandem MS
Column: UCT Selectra® DA LC column, 100 × 2.1 mm, 1.8 µm
Guard Column: UCT Selectra® DA guard column, 10 × 2.1 mm, 1.8 µm
Injection Volume: 10 µL
Mobile Phase A: 0.1% formic acid in D.I. H2O
Mobile Phase B: 0.1% formic acid in methanol
Column Flow rate: 0.40 mL/min
Results
Table I: Accuracy and precision data in plasma using free hormone procedure (n=6 each).
Table II: Accuracy and precision data in plasma using total hormone procedure (n=6 each).
Conclusion
Excellent recoveries ranging from 87.1% to 99.8% were obtained using UCT's mixed-mode DAU extraction column (CSDAU203) and procedure in conjunction with the Selectra® DA UHPLC column (SLDA100ID21-18UM) fothe analysis of free and total thyroid hormones in human serum and plasma.
UCT, LLC
2731 Bartram Road, Bristol, PA 19007
tel. (800) 385-3153; Email:
Website:
Articles in this issue
about 10 years ago
Vol 34 No 2 LCGC North America February 2016 Application Notebook PDFabout 10 years ago
Semivolatiles Analysis Using Split Injectionabout 10 years ago
Ethylcarbamate (Urethane) in Distilled Alcoholic Beveragesabout 10 years ago
Flexible Anion Analysis with Hamilton PRP-X100 and X110 ColumnsNewsletter
Join the global community of analytical scientists who trust LCGC for insights on the latest techniques, trends, and expert solutions in chromatography.