10 common GC mistakes - are you guilty of making them?
1) Not Taking Enough Care at the Inlet
Any leak in GC will cause significant problems with chromatography and the inlet is the most likely area. Don't take the risk of having to troubleshoot a leak, change the septum regularly and make sure the column is correctly installed and leak-free.
2) Turning the oven on before the carrier gas
If there's one thing that will permanently damage a column more quickly than anything else it's heating the column without any carrier gas flowing. Get into the habit of making sure the carrier gas is flowing before you switch on the oven AND make sure the oven is cool before switching the carrier off. When switching the oven off, set the oven temperature to 35oC before switching the oven off as this will cool much quicker.
3) Not programming in column dimensions
Most modern GCs use an EPC (electronic pneumatic controller) to accurately deliver carrier flows and this relies totally on the correct column dimensions being entered by the operator. Make sure anytime you change the column you enter the correct dimensions into the GC or else your flowrate and split ratios will be wildly inaccurate.
4) Not using optimal flowrate for carrier
Each carrier gas has an optimal linear velocity, which gives the best efficiency (Nitrogen c.12cm/sec, Helium c.35cm/sec and Hydrogen 40 - 50cm/sec). To obtain this optimal linear velocity the flowrate is matched to the internal diameter of the column in use. e.g. for a 0.32mm ID column using helium the optimal flowrate will be about 2.0ml/min. If the optimal flowrate is not used, column efficiency can be significantly reduced.
The complete article is available to CHROMacademy Lite and Premier members here >>
CHROMacademy Lite Membership is FREE and it only takes two minutes to register.
With a Lite Membership you are given access to:
∙ This month's webcast & tutorial
∙ Selected eLearning modules
∙ Featured CHROMacademy Content
∙ The CHROMacademy forum
Test drive CHROMacademy and Check out more great content available FREE to our Lite members »
The Reality Behind Column Insertion Distance
September 10th 2024Column insertion distance is critical to good chromatography. What happens if the column is installed too low in the injection port? Is insertion distance more important when performing split injection or splitless injection? Does the position of the column in the injection port impact reproducibility?
Modern HPLC Strategies: Improving Retention and Peak Shape for Basic Analytes
August 16th 2024In high-performance liquid chromatography (HPLC), it is common for bases and unreacted ionized silanols on silica-based columns to cause irreproducible retention, broad peaks, and peak tailing when working with basic analytes. David S. Bell, Lead Consultant at ASKkPrime LLC offers innovative HPLC strategies that can help mitigate such issues.