The Application Notebook
The Agilent 1290 Infinity LC has significant capabilities for a wide range of HPLC and UHPLC applications. It exhibits a broader power range (that is, the combination of pressure and flow capabilities) than any other commercially available system and has the flexibility to operate with a wide range of column dimensions and particle sizes. Additionally, advanced optical design in the diode array detector allows a wide dynamic range and high sensitivity, both of which are critical in the monitoring of small impurities in fine chemicals.
Michael Woodman, Agilent Technologies Inc., Chemical Analysis Solutions, Wilmington, Delaware, USA.
The Agilent 1290 Infinity LC has significant capabilities for a wide range of HPLC and UHPLC applications. It exhibits a broader power range (that is, the combination of pressure and flow capabilities) than any other commercially available system and has the flexibility to operate with a wide range of column dimensions and particle sizes. Additionally, advanced optical design in the diode array detector allows a wide dynamic range and high sensitivity, both of which are critical in the monitoring of small impurities in fine chemicals.
The combined benefits are demonstrated by a separation of impurities found in a sample of octyl-dimethyl-4-aminobenzoate. The high pressure capability of the system allows the use of methanol, as well as acetonitrile, to explore the selectivity of the two solvents. At 1.5 mL/min, using a simple 2-min gradient and a 3.0 mm × 50 mm 1.8 μm column, the analysis time is only 3 min. The separation of the main components is shown in Figure 1.
Figure 1: Initial separation conditions showing a need for greater resolution and selectivity. Sample: Octyl dimethyl para-aminobenzoate, 1 mg/mL. Gradient: 1.5 mL/min, 40% to 90% ACN/water over 2 minutes. Up to 460 bar on ZORBAX StableBond RRHT C18, 3 mm à 50 mm, 1.8 μm, 40 °C. 0.75 minute retention: 4-amino-benzoic acid; 2.1 minute retention: Octyl dimethyl para-aminobenzoate.
The speed, resolution and flexibility of the system are further demonstrated by a separation of the sample using methanol or acetonitrile with low solvent consumption 2.1 mm i.d., 1.8 μm columns. The flow rate and gradient conditions are optimized for each solvent, to produce a gradient separation with maximum pressure of approximately 850 bar, a conservative setting for the 1200-bar capability of the Agilent 1290 Infinity LC. The separation of the main components, with the two organic solvents, is shown in Figure 2(a) (acetonitrile, top panel) and 2(b) (methanol, lower panel), where the chromatograms are zoomed to the region of peaks shown from approximately 1.2–2.5 min in Figure 1.
Figure 2: Results using ACN vs MeOH with the same gradient slope on the 1290 Infinity LC. Sample: ODPABA working standard, 1 mg/mL. Conditions: ACN gradient 0.6 mL/min, 40% to 90% ACN/water over 7.4 minutes. Up to 850 bar on ZORBAX StableBond RRHD C18 2.1 mm à 150 mm, 1.8 μm, 40 °C. Methanol gradient 0.52 mL/min, 50% to 100% MeOH/water over 5.7 minutes. Up to 850 bar on ZORBAX StableBond RRHD C18 2.1 mm à 100 mm 1.8 μm, 40 °C. The increased selectivity of methanol allowed a shorter column to be used, decreasing run time and avoiding the use of more expensive acetonitrile mobile phase.
The combined high flow and high pressure capability of the system allows one to use high efficiency columns, producing rapid separations with remarkable resolution while conserving solvent over the use of 4.6 mm i.d. columns. Impurity detection, as a result of high detector sensitivity and stability, is estimated to be < 0.01%.
Agilent Technologies Inc.,
2850 Centerville Road, Wilmington, Delaware 19808, USA
tel. +1 800 227 9770 fax +1 302 633 7801
E-mail: info_agilent@agilent.com Website: www.agilent.com/chem
HPLC 2025 Preview: The Present and Future of Automation in Analytical Laboratories
May 22nd 2025Analytical laboratories are undergoing a fundamental transformation. In the face of increasing sample volumes, growing regulatory requirements, and the rising demand for faster, more precise, and cost-efficient analysis, optimizing laboratory processes is becoming a central focus. Automation technologies offer promising solutions in this regard. Recently, they have evolved from isolated solutions to comprehensive systems that permeate nearly all areas of laboratory practice. This development not only opens up new opportunities in terms of efficiency, data quality, and scalability but also brings technical, organizational, and personnel challenges. To successfully address these, strategic approaches are needed that consider both the technological and human dimensions of the transformation.
Quantifying Isavuconazole in Dried Blood Spots Using HPLC
May 21st 2025Isavuconazole, an antimycotic agent used to treat fungal infections, can typically be found during dried blood spot sampling. However, there are obstacles that keep it from being an ideal approach for properly determining the drug’s presence.
Separating and Quantifying Spinach Flavonoids with UHPLC–MS/MS
May 21st 2025Researchers at Baylor College of Medicine’s Children's Nutrition Research Center (Houston, Texas) developed and validated a high-throughput extraction and ultrahigh-performance liquid chromatography–tandem mass spectrometry (UHPLC–MS/MS) method to separate and quantify 39 spinach flavonoid species in 11.5 min.