News|Articles|September 16, 2025

Building a reliable method for tirzepatide

This step-by-step method development study for a GLP-1 receptor agonist provides a robust and precise analytical workflow for purity testing of tirzepatide.


Related to this article

Chromatography Detects Toxins in Tea Samples
High-performance liquid chromatography-tandem mass spectroscopy (HPLC–MS/MS) chromatography reveals pyrrolizidine alkaloids (PA)/pyrrolizidine alkaloid N-oxides (PANO) toxin risks in 63 tea samples.
Figure 2. Split injection used to measure triplicate test of MTBE, hexane, o-xylene, and 1-methylnaphthalene comparing different solvents (methanol, methylene chloride) and column insertion distances starting with 0.5 mm, 5.0 mm, 10.0 mm, 15 mm, and 20.0 mm. Methanol had the best performance at the 5.0 mm insertion distance and methylene chloride looked slightly better at the 0.5 mm. We would still recommend not going below the 5.0 mm manufacturer recommended insertion distance.
Revisiting insertion distance for methanol and methylene chloride in GC analysis and confirming that insertion depth strongly affects response and reproducibility.