Implications of a Stronger Dependence on Solvent Composition
The fact that large molecules exhibit a stronger dependence of retention on mobile phase composition has been known for decades (10). When reviewing retention data for molecules covering a large range in molecular weight, Snyder and Dolan found that S can be estimated from the molecular weight (MW) using equation 8 (this assumes we are working with the natural logarithm of retention factor, as in Figure 1, and the MW is in Daltons):
Nevertheless, I think it is instructive to look at real data, and the implications that flow from this big diƒerence in behavior between small and large molecules. Figure 1 shows the dependence of retention factor (logarithmic scale) on mobile phase composition for three molecules of diƒerent size in the range of about 100 to 6000 Da. When these data are fit to a linear function of the mobile phase composition, as in LSST, the slopes are -7.49, -10.6, and -268 for the three molecules in order of increasing size. Using the slope and intercept from this plot, we can then calculate the local retention factor of the analyte as it travels from the inlet to the outlet of the column. As discussed above, the local retention factor of the analyte as it exits the column (ke) determines the peak width as dictated by equation 5. In this special case, the klocal at 50 mm from the inlet in Figure 2 is equal to the ke in equation 5. In Figure 2 we see a dramatic difference between the two smaller molecules and the 6 kDa biomolecule. For the two smaller molecules, we see that the local retention factor is about 2 when they exit the column, whereas the local retention factor of the larger molecule is near zero. Since the peak width dictated by equation 5 is proportional to the factor ke+1, this means that we would expect the peak for the biomolecule to be about one-third that of the smaller molecules, provided that the plate number for these three molecules is about the same (Note: this is an important caveat that must be assessed before making comparisons of experimental peak widths). On one hand, this difference in behavior is highly beneficial for separations of large molecules because it means that we have the potential to observe very narrow peaks under gradient elution conditions. On the other hand, however, this also means that these separations are prone to serious negative effects of extra-column dispersion that can make the peaks broader that can make the peaks broader than they should be–not because the column is not performing well, but because the narrow peaks exiting the column are compromised by broadening on the way to, and through, the detector. Readers unfamiliar with the details associated with these extra-column eƒects are referred to a series of recent articles in this magazine that focused on their origins, avoidance strategies, and a free tool that can be used to estimate their eƒects on real separations (11–14). A note of caution is warranted here. The ke values for large molecule depend very strongly on gradient slope. Thus, it is not always the case that very narrow peaks should be observed for large molecules. Readers are encouraged to use any of a number of free tools (for example, www.multidlc.org/ hplcsim; www.multidlc.org/MultiSimLC) to explore the effects of variables such as column length, gradient range (that is, the range of mobile phase composition covered in the gradient), and flow rate on the expected peak widths under conditions relevant to their experiments to develop a sense for what kind of peak widths to expect.
The discussion in this section has implicitly assumed nominally ideal behavior of the large molecule—that is, no overload, no stickiness, or other bad behavior. This ideal case serves as an important baseline for setting our expectations about peak width, but we also have to realize that there are many ways for things to go badly with large molecules that contribute to peak width in ways not accounted for by the preceding discussion.
Effect of Pore Size
Most stationary phase particles designed for small molecule separations have average pore diameters in the neighborhood of 80 to 120 Å. Small molecules typically have effective diameters less than about 10 Å, so they can diƒuse into and out of the pores of the particle with relatively little obstruction. As the molecular weight increases, however, the diffusion of the analyte through the pores becomes more obstructed, leading to significant peak broadening particularly at high flow rates. The continuing growth of interest in high performance separations of biomolecules has motivated the commercialization of new stationary phase materials with pore sizes exceeding 500 Å, and the benefits of such materials for separations of large biomolecules are compelling (16).
Volume Overload
Volume overload refers to a situation where the volume of the injected sample, and quite often the composition of the sample solvent, affects the shapes and widths of peaks for analytes of interest. Sometimes, this is difficult to avoid; the sample may be a portion of a reaction mixture that contains a high level of organic solvent, for example. We have illustrated how troublesome this can be, and some potential solutions, in prior installments of this column (17,18). However, the point I’d like to make here is that both the degree of the problem, and the ease with which it can be fixed, are usually exaggerated with large molecules. This, again, is because of the stronger dependence of retention on solvent composition for large molecules as illustrated in Figure 1. For example, a mixture of large molecules might be separated using a gradient that only runs from 10 to 20% acetonitrile (ACN). In this case, a sample that contains just 5% ACN more than the starting point in the gradient can lead to disastrous chromatography (19). The good news, however, is that the strong solvent dependence of retention can also be the source of an easy fix. In my example here, simply diluting the sample 1:1 with water would decrease the ACN level to 7.5%, and the problem would entirely disappear.
Strongly Adsorbing Analytes–or "Stickiness"
Large molecules can exhibit particularly strong adsorption to column materials and LC system components, leading to widerthan-expected peaks. In some cases this is because of a large number of functional groups for each analyte molecule that can set up a Velcro-like sticking situation. For example, phosphates are known to adsorb strongly to stainless steel. When an analyte such as an oligonucleotide with many phosphate groups on one molecule encounters a steel surface, it can stick nearly irreversibly unless the surface is passivated with phosphoric acid first, or an additive is added to the mobile phase to block these problematic interactions. Readers interested in learning more about this problem and potential solutions are referred to prior “LC Troubleshooting” installments that address the topic in more detail (20–22).
What to Do When the Peak Widths Are Not as Expected
In my laboratory, we consider the following possibilities when we suspect that a column is producing peaks that are wider than expected. Each of these bullet points was discussed in a bit more detail in Part VI of this series. Readers interested in learning more about these potential troubleshooting paths are referred to the previous installment for a more thorough discussion.
- Do a health check on the column
- Consider the possibility that wide peaks are due to chemical problems
- Check to make sure that the system is not the problem
Summary
In this seventh installment on essential topics in LC troubleshooting, I have highlighted additional considerations that are important to keep in mind when estimating expected peak widths for separations of large molecules. Large molecules exhibit a stronger dependence of retention on mobile phase composition, which aƒ ects not only how narrow the peaks can be relative to peaks for small molecules, but also the sensitivity of separation quality to other factors such as volume overload. Additionally, there are other factors aƒ ecting peak width that are particularly important with large molecules, including the eƒ ect of stationary phase pore size, and functional groups that can lead to very strong analyte adsorption. Developing both expectations about peak widths for large molecules when they behave ideally, and an understanding about additional factors that can lead to non-ideal behavior, can be powerful tools when troubleshooting underperforming large molecule separations.
References
(1) Stoll, D. R. Essentials of LC Troubleshooting, Part I: Pressure Problems. LCGC N. Am. 2021, 39 (12), 572–574.
(2) LCGC Troubleshooting Wallchart, (n.d.). https://www.chromatographyonline.com/view/troubleshooting-wallchart
(3) Stoll, D. R. Essentials of LC Troubleshooting, Part II: Misbehaving Retention Times. LCGC N. Am. 2022, 40 (4), 151–155. DOI: 10.56530/lcgc.na.fa1867h1
(4) Cash, K.; Seid, E.; Dahlseid, T.; Stoll, D. R. Essentials of LC Troubleshooting, Part II, Continued: More Misbehaving Retention Times. LCGC N. Am. 2023, 41 (3), 90–94. DOI: 10.56530/lcgc.na.ja4681f2
(5) Stoll, D. R. Essentials of LC Troubleshooting, III: Those Peaks Don’t Look Right. LCGC N. Am. 2022, 40 (6), 244–247. DOI: 10.56530/lcgc.na.az3781g7
(6) Stoll, D. R. Essentials of LC Troubleshooting, Part IV: What Is Going On with the Baseline? LCGC N. Am. 2022, 40 (8), 338–342. DOI: 10.56530/lcgc.na.tk9778b8
(7) Stoll, D. R. Essentials of LC Troubleshooting, Part V: What Happened to My Sensitivity? LCGC N. Am. 2022, 40 (10), 471–475, 483. DOI: 10.56530/lcgc.na.uk2876b9
(8) Stoll, D. R. Essentials of LC Troubleshooting, VI: How Wide Should Those Peaks Be? LCGC N. Am. 2022, 40 (12), 562–565, 578. DOI: 10.56530/lcgc.na.cb1781m7
(9) Carr, P. W.: Stoll, D. R. Speed and Performance in Liquid Chromatography, in Multi-Dimensional Liquid Chromatography: Principles, Practice, and Applications; CRC Press, 2022.
(10) Snyder, L. R.; Dolan, J. W. Separation of Large Molecules, in High-Performance Gradient Elution: The Practical Application of the Linear-Solvent-Strength Model, John Wiley, 2007: pp. 228–282.
(11) Stoll, D. R.; Broeckhoven, K. Where Has My Efficiency Gone? Impacts of Extracolumn Peak Broadening on Performance, Part I: Basic Concepts, LCGC N. Am. 2021, 39 (4), 159–166.
(12) Stoll, D. R.; Broeckhoven, K. Where Has My Efficiency Gone? Impacts of Extracolumn Peak Broadening on Performance, Part II: Sample Injection. LCGC N. Am. 2021, 39 (5), 208–213.
(13) Stoll, D. R.; Broeckhoven, K. Where Has My Efficiency Gone? Impacts of Extracolumn Peak Broadening on Performance, Part III: Tubing and Detectors. LCGC N. Am. 2021, 39 (6), 252–257.
(14) Stoll, D. R.; Lauer, T. J.; Broeckhoven, K. Where Has My Efficiency Gone? Impacts of Extracolumn Peak Broadening on Performance, Part IV: Gradient Elution, Flow Splitting, and a Holistic View. LCGC N. Am. 2021, 39 (7), 308–314.
(15) Gilar, M.; Hill, J.; McDonald, T. S.; Gritti, F. Utility of Linear and Nonlinear Models for Retention Prediction in Liquid Chromatography. J. Chromatogr. A 2019, 460690. DOI: 10.1016/j. chroma.2019.460690
(16) Kirkland, J. J.; Schuster, S. A.; Johnson, W. L.; Boyes, B. E. FusedCore Particle Technology in High-Performance Liquid Chromatography: An Overview. J. Pharm. Anal. 2013, 3, 303–312. DOI: 10.1016/j.jpha.2013.02.005
(17) Stoll, D. E. Mixing and Mixers in Liquid Chromatography, Part III: Solutions for Problems with Sample Diluents. LCGC N. Am. 2019, 37 (6), 368–373.
(18) Stoll, D. R. What’s Trending in LC Troubleshooting? LCGC N. Am. 2019, 32 (1), 18–23.
(19) Stoll, D. R.; Lhotka, H. R.; Harmes, D. C.; Madigan, B.; Hsiao, J. J.; Staples, G. O. High Resolution Two-Dimensional Liquid Chromatography Coupled with Mass Spectrometry for Robust and Sensitive Characterization of Therapeutic Antibodies at the Peptide Level. J. Chromatogr. B 2019, 1134–1135, 121832. DOI: 10.1016/j.jchromb.2019.121832
(20) Hsiao, J. J.; Staples, G. O.; Stoll, D. R., Troubleshooting LC Separations of Biomolecules, Part I: Background, and the Meaning of Inertness. LCGC N. Am. 2020, 38 (3), 146–150.
(21) Hsiao, J. J.; Chu, T.-W.; Potter, O. G.; Staples, G. O.; Stoll, D. R., Troubleshooting LC Separations of Biomolecules, Part II: Passivation and Mobile-Phase Additives. LCGC N. Am. 2020, 38 (8), 431–434.
(22) Stoll, D. R. Strongly Adsorbing Analytes: What, Why, and How to Fix It. LCGC N. Am. 2023, 41 (7), 242–244. DOI: 10.56530/lcgc.na.vr7280f7
About the Author