For many years, the use of guard columns has been advocated by column manufacturers and other experts to protect and extend column lifetime and performance of analytical columns from potential damage caused by the presence of chemical contaminants and microparticulates in the sample and mobile phase.
Philip J. Koerner, J. T. Presley, and Tom Cleveland, Phenomenex, Inc.
For many years, the use of guard columns has been advocated by column manufacturers and other experts to protect and extend column lifetime and performance of analytical columns from potential damage caused by the presence of chemical contaminants and microparticulates in the sample and mobile phase. The introduction and adoption of smaller particle size (< 3 μm) and narrow ID columns (≤ 2 mm) columns for ultra-high performance liquid chromatography (UHPLC) presents additional challenges. Specifically, these ultra-high performance columns are very sensitive even to small increases in dead volume and the use of traditional guard columns or guard cartridge systems will show a marked decrease in column performance (efficiency and resolution). However, small particle columns are even more susceptible to contamination and blockage from microparticulates that may be present in both the sample and mobile phases as they can occlude the smaller porosity column frits and dramatically shorten column lifetime.
KrudKatcher Ultra in-line filters are specifically designed for use with ultra-high performance columns from 1.0 to 4.6 mm ID and are pressure rated to 20,000 psi (1,375 bar). The filter body houses an integrated 0.5 μm 316 stainless steel depth filter that efficiently removes microparticulates from the flow stream without contributing to system backpressure or dead volume (<0.2 μL).
To illustrate the benefit of using the KrudKatcher Ultra in-line filter for extending column lifetime, a chromatographic separation of a ketone mixture on a Kinetex® 2.6 μm PFP 50 × 2.1 mm column was set up on a UHPLC system (Figure 1). The initial injection shows excellent peak shape and was maintained for over 1,100 injections (1a). However, after 1,175 injections the chromatographic performance has noticeably degraded resulting in split peaks with some peak broadening and tailing evident (1b). Simply replacing the KrudKatcher Ultra and re-injecting the sample returns the original chromatographic performance (1c). The KrudKatcher Ultra in-line filter successfully protected the ultra-high performance Kinetex column from damaging microparticulates in the mobile phase flow path and sample that would have dramatically shortened the column lifetime.
Figure 1
The regular use of KrudKatcher™ Ultra in-line filters with ultra-high performance UHPLC columns protects and extends the lifetime by protecting them from damaging microparticulates that may be present in the sample and mobile phase. Additionally, the KrudKatcher Ultra does not have a detrimental effect on chromatographic performance (retention time, peak width, resolution, or efficiency). For additional information, please request technote TN-1063.
Phenomenex, Inc.
411 Madrid Ave., Torrance, CA 90501
tel. (310)212-0555, (310)328-7768
Website: www.phenomenex.com
Inside the Laboratory: Measuring Molecular Tracers of Emerging Aerosol Sources
October 14th 2024Inside the Laboratory is a joint series with LCGC International and Spectroscopy, profiling analytical scientists and their research groups at universities all over the world. This series spotlights the current chromatographic and spectroscopic research their group is conducting, and the importance of their research in analytical chemistry and specific industries.
Top-down characterization of engineered Bcl-xL proteoforms
October 11th 2024Top-down fragmentation enables rapid characterization of phosphorylated proteins without extensive sample preparation and digestion. In this study, electron capture dissociation (ECD) was used to fragment proteoforms of the cell death-related protein, Bcl-xL. Using these methods, 85–90% sequence coverage was achieved for Bcl-xL proteoforms, allowing for effective localization of phosphorylation within minutes.
Antibody peptide mapping using the new Agilent ExD cell
October 11th 2024Enhanced antibody analysis using electron capture dissociation (ECD) allows for precise glycan localization in low-abundance glycopeptides. This study compares the fragmentation of trastuzumab tryptic digests using ECD and collision-induced dissociation (CID). While CID generates abundant glycan HexNAc ions at 204 m/z, ECD preserves the labile glycan group, enabling accurate site localization.