Waters Application Note
Fast, flexible platforms for peptide quantification are needed, particularly for a discovery setting. This type of methodology would be especially advantageous in the case of amyloid beta (aβ) peptides. The deposition/formation of insoluble aggregates, or plaques, of aβ peptides in the brain is considered to be a critical event in the progression of Alzheimer's disease (AD) and thus has the attention of many researchers. A previous Waters application note (720003682en) described in detail the development of a fast, flexible SPE-LC–MS–MS platform for the quantification of multiple aβ peptides from human or monkey CSF for use in a biomarker or preclinical discovery setting. In this work, the mass spectrometry platform has been updated from the Xevo TQ MS to the Xevo TQ-S mass spectrometry system. This change facilitated both a 4× reduction in required sample size and a 4–5× increase in assay sensitivity. This work focuses on methods for the 1-38, 1-40 and 1-42 aβ, Table 1.
Table 1: Sequence, MW and pI information for amyloid peptides.
SPE-LC–MS–MS Conditions
LC system: Waters ACQUITY UPLC System
Column: ACQUITY UPLC BEH C18 300 Å, 2.1 × 150 mm, 1.7 µm, Peptide Separation Technology
SPE: Oasis MCX µElution 96-well plate, 50 µL human or animal CSF
MS system: Waters Xevo TQ-S, ESI+
Figure 1: Representative ESI+ MS-MS spectrum for amyloid 1-42 with fragment sequence ions labelled.
• An improved SPE-UPLC–MS–MS bioanalytical method was developed and validated for the simultaneous quantification of multiple amyloid β peptides in human CSF.
• MS was performed in positive ion mode since CID of the 4+ precursor ion yielded several distinct product ions corresponding to inherently specific b sequence ions (representative spectrum shown in Figure 1).
• UPLC separation of the three amyloid β peptides is shown in Figure 2.
Figure 2: Representative UPLCâMSâMS analysis of amyloid 1-38, 1-40 and 1-42 peptides extracted from artificial CSF + 5% rat plasma.
• The increased sensitivity of the Xevo TQ-S triple quadrupole mass spectrometer facilitated the use of 4× less sample and a 4–5× improvement in quantification limits (Table 2).
Table 2: Comparison of Standard Curve and QC range using Xevo TQ and TQ-S MS.
• Average basal levels and RSD values for all 3 aβ peptides in 2 sources of human CSF are shown in Table 3 and are lower or equal to 5%.
Table 3: Baseline levels of amyloid peptides in 2 sources of pooled human CSF.
• Overspiked QC samples were prepared in triplicate in 2 sources of pooled human CSF at 0.04, 0.075, 0.15, 0.2, 0.8, 2 and 6 ng/mL. Accuracy and precision values met the regulatory criteria for LC–MS–MS assays. Results from QC sample analysis are shown in Table 4. Average deviation from expected is 2.3%
The method described herein eliminates time-consuming immunoassays or immunoprecipitation steps for pre-clinical work.
Table 4: Average deviation values for all overspike QC samples
The use of a single UPLC–MS–MS assay represents a significant advantage over an ELISA assay, which would require multiple assays with multiple antibodies to quantify each of the relevant peptides.
Copyright 2010 : ACQUITY UPLC, Oasis, The Science of What's Possible, Xevo are trademark of Waters Corporation.
Waters Corporation
34 Maple Street, Milford, Massachusetts 01757, USA
tel. +1 508 478 2000 fax +1 508 478 1990
Website: www.waters.com
Measuring Procyanidin Concentration in Wines Using UHPLC
January 24th 2025Researchers from the University of Bordeaux (Villenave d'Ornon, France) report the development and validation of a rapid and quantitative analytical method measuring crown procyanidin concentration in red and white wines using ultra-high performance liquid chromatography (UHPLC) coupled with a ultra-high performance liquid chromatography (Q-TOF) mass spectrometer.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ions used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.