
- LCGC Europe-04-01-2010
- Volume 23
- Issue 4
Investigating Solvent Purity Using Comprehensive Gas Chromatography: A Study of Acetones
This work examines the impurities present in a variety of acetone sources using comprehensive gas chromatography.
Broad spectrum chemical analysis of trace level components is a continuing challenge for any analytical chemist. This challenge is further confounded when chemical impurities may be present in common organic solvents or when chemical artifacts may be formed, produced and introduced during an analytical procedure. Minimizing and understanding these chemical artifacts is critical for trace level detection and is crucial for accurate analytical conclusions. Comprehensive gas chromatography is an excellent analytical tool to help address these complex mixture challenges. This work examines the impurities present in a variety of acetone sources using comprehensive gas chromatography. This work highlights the extreme variability possible in solvent sources and hence the importance of understanding the impurities that may confound an analytical method or result.
General research or specific experimental protocols involving trace and ultra-trace chemical analysis and detection is an extremely challenging and daunting task for scientists. As well as the inherent chemical background that is present when a sample is collected, numerous chemical analyses and methods rely on a pure solvent for experimental procedures. For example, many chemical extraction approaches use a solvent to extract target chemicals of interest from a substrate (e.g., soil) or collection sorbent. After the extraction step, an evaporation/condensation/concentrating step is performed to reduce the extraction volume to improve detection of trace components. Unfortunately, this volume reduction step will also concentrate many of the impurities present within a solvent and can further confound research efforts. Many of the solvent artifacts introduced into a sample are from chemical impurities that are not completely removed by standard manufacturing processes or techniques for the solvent, as well as the glassware and storage vials used. Regardless of the source, minimizing and understanding these chemical artifacts is critical for trace level detection and is crucial for unambiguous analytical results.
Acetone is one of the most widely used solvents in the laboratory environment. A common chemical production method for acetone is through the cumene process.1 In the cumene process benzene is alkylated with propene resulting in cumene (isopropylbenzene), which is subsequently oxidized to give phenol and acetone. A wide variety of impurities may exist in the final crude acetone fraction (e.g., aliphatic aldehydes, olefins and carbonyl impurities, and more specifically diacetone alcohol, mesityl oxide and hydroxyacetone), which is subsequently purified to obtain the quoted purity. Once the acetone has been obtained in the laboratory, further use or storage may form additional chemical artifacts. Hunchak and Suffet2 and Lafleur and Pangaro3 examined acetone–hexane artifacts produced during a soxhlet extraction process. The chemical artifacts produced through the various procedures and reactions include diacetone alcohol and mesityl oxide as well as phorone, mesitylene and isophorone.2,3
The work presented here examines the purity of various acetone sources using comprehensive gas chromatography and mass spectrometry (GC×GC–MS) for broad spectrum chemical analysis. The purpose of this work is to characterize the impurities present in a solvent as a possible approach to track or match acetone sources for chemical forensic investigations. In particular, in the preparation of the toxin ricin, acetone is a common solvent used for the extraction of the castor oil. Previous work developed a simple solid-phase microextraction (SPME) approach to detect the use of acetone as an extraction solvent in the final crude ricin preparation.4 The motivation for this study is to provide additional sample processing information regarding the type/grade of acetone used in the production of ricin or other extracted materials. As broad spectrum chemical analysis becomes increasingly important GC×GC–MS has emerged as a powerful analytical tool. Comprehensive GC×GC is a multidimensional separation technique that separates a sample sequentially on two different columns, which are chosen to separate mixtures based on distinct — ideally orthogonal — chemical properties. Multidimensional separations have the advantage of greatly improved separations compared to conventional single-dimension approaches.5–9 The work presented here is the first step, examining the impurities present in various acetone sources. However, the overall importance of understanding the possible impurities present in a solvent is vital and can be applied to any trace analytical method.
KEY POINTS
Experimental Section
Chemicals
Table 1 lists the 21 acetone sources obtained and examined for this study. The mention of trade names or commercial products in this work does not constitute an endorsement, criticism or recommendation for their use.
Table 1: Acetone source list.
Chemical Analysis
Chemical analyses were performed on a Leco Pegasus 4D GC×GC–MS system (LECO, St Joseph, Michigan, USA) equipped with a Gerstel cooled injection system (CIS4) and multipurpose sampler (MPS2) (Gerstel, Baltimore, Maryland, USA). The separation used ultra-high purity helium as the carrier gas, set at 1 mL/min, constant flow. Because the type and number of impurities were largely unknown for each acetone source, two different comprehensive GC×GC column sets were used (see Table 2, A×C and B×C) for all 21 acetone sources. The first dimensional column was either a polar or non-polar type and the secondary column was of moderate polarity and remained constant throughout the study. Unless stated otherwise the experimental conditions are listed in Table 2. LECO ChromaTOF software (version 3.32) was used for data collection and analysis. Higher level data analysis was performed using an in-house software tool (Data Analysis Tool Extension, DAnTE,
Table 2: Experimental conditions for chemical analysis.
Results and Discussion
In this study 21 different acetone sources (brands) were analysed to examine the trace impurities inherently present within the neat acetone source. Table 1 lists the 21 acetone sources, which range from high purity laboratory grade to hardware store grade to commercial finger nail polish remover. Within this acetone sample set the supply containers included dark amber glass, clear glass, plastic and metal. All acetone sources were quoted above 99% purity. In comparison to the extraction solvent artifacts referenced by Hunchak and Lafleur,2,3 this study attempts to examine the impurities inherently present within an acetone source. To minimize acetone-related artifacts during the analysis, a cooled injection approach is used where 5 µL of acetone is slowly injected into a cooled injection system (CIS) with the majority of injected acetone vented during this process (see Table 2 for experimental details). Following the acetone injection process into the CIS, the solvent vent is closed and the CIS is rapidly heated to inject the remaining collected residual impurities.
To offer a baseline perspective of the purity of a low grade acetone source, Figure 1 shows a comparison between a 1 µL split injection of moderate split ratio (50:1) and 5 µL CIS injection of acetone A11 (hardware store brand) obtained using column set A×C. In these and the following 2D chromatograms, the blue color represents the baseline signal and the red represents higher signal values of compounds. As can be seen in Figure 1, at this 1 µL split injection level (top chromatogram) a number of trace compounds (peaks) are visually apparent, but with acetone the dominant compound in this separation. The major impurity compounds are listed in Figure 1 and include 2-pentanone, 4-hydroxy-4-methyl- (diacetone alcohol) and toluene. Pentane is observed as carryover from the syringe wash solvent used between injections. The horizontal solute bands at the end of the chromatogram are due to column bleed components. In general even this low grade acetone source appears to be relatively pure at this injection level. However, when a 5 µL CIS injection of the same A11 is examined the number of impurities observed dramatically increases as shown in the lower chromatogram in Figure 1 where the vertical dashed line denotes the start of the data acquisition for the larger injection volume. These impurity components encompass a wide range of polarity, volatility and analytical stability. Depending upon the source and type of acetone, numerous trace impurities are found in all 21 acetone sources and all the sources can easily be differentiated by visual inspection of the chromatograms.
Figure 1
Figure 2 summarizes the separations obtained for five of the acetone sources (A3, A9, A20, A23 and A27) for both column sets (i.e., A×C and B×C) acquired using the 5 µL CIS injection approach. A number of observations are illustrated in Figure 2. In general, among all the acetone sources the number and amount of impurities can be extremely large and variable. The general chemical classes observed include saturated and unsaturated hydrocarbons, organic acids, ketones and branched aromatic hydrocarbons according to their chromatographic elution order and mass spectral information. Interestingly, regardless of the column set used (i.e., A×C or B×C) various solute zones appear as large streaks typically early in the 2D chromatograms, which was also observed in Figure 1. This solute zone streaking is probably due to the chemical mismatch between the solute and stationary phase (e.g., non-polar solute and a polar stationary phase and vice versa) of the 1st dimensional column so that there is a lack of adequate solute focusing during the injection process. For example, in acetones A3 and A9, both acetic acid and propanoic acid are found by using column set A×C with good solute zone shape; however, acetic acid is absent within the data collection window used and propanoic acid appears as a large streak when using column set B×C (Figure 2). Obviously the presence of these trace acids in the solvent could have deleterious effects for some solutes (e.g., reactions producing artifactual solutes) and could confound trace level applications. Importantly, these organic acids could easily go undetected if only a non-polar type column were used. Conversely, many of the other acetone sources have a large number of what appear to be hydrocarbon type impurities according to the mass spectral information. These hydrocarbon type compounds appear as streaks early in the 2D chromatograms when using column set A×C (e.g., acetone sources A9 and A27 in Figure 2, as well as A11 in Figure 1 as highlighted by black outlined regions), but appear as narrow solute zones with column set B×C. This hydrocarbon source region is outlined for acetone source A9 in Figure 2 and contains numerous straight chain, branched and cyclo hydrocarbon compounds as indicated by their electron impact mass spectra. Varying amounts of these hydrocarbons can be found in all the acetone sources examined and the solute zone shape is more clearly defined when using column set B×C. Again, these solutes may go undetected if using column set A×C due to the lack of solute focusing/streaking. This result highlights the importance of using a number of different primary columns of different chemical nature when examining for trace level components in a broad spectrum chemical analysis. (Note: As an aside, when using a common 2nd dimensional column and changing only the 1st dimension column the down time of the GC×GC–MS instrument is minimal, as the mass spectrometer may not need to be vented during this column switch.)
Figure 2
The hydrocarbon region outlined in A9 in Figure 2 is examined more closely in Figure 3 with a comparison between acetone sources A9 and A27. In comprehensive 2D separations compared to conventional 1D GC separations, chemically-similar compounds will create patterns or groupings of solute zones (e.g., a homologous series of saturated hydrocarbons) within the 2D separation plane. These chemical groupings or families in a 2D chromatogram can be of help in identifying unknowns, when no standard components are available, in the absence of commercial MS library spectra or when the resulting experimental MS spectra are very similar to others. In Figure 3, three m/z values are selected to highlight groupings of saturated hydrocarbons (m/z = 57), olefins (m/z = 69) and alkyl substituted benzene rings (m/z = 91), which also captures possible mesitylene, cumene and their isomeric impurities. In general, as expected, the A9 acetone source appears to be of higher chemical purity than A27 as depicted by the smaller number of solute zones. Interestingly, acetone source A9 also appears to have a greater fraction of saturated and unsaturated hydrocarbon compounds that elute later in time. This implies that A9 has a larger fraction of higher boiling point hydrocarbons than does A27, which may indicate a certain type or degree of purification process. Finally, a series/family of alkyl substituted benzyl compounds (m/z = 91) are observed in both acetone sources, but the amounts present in acetone A27 are much larger. Again the presence of these benzyl compounds may indicate the type or extent of the purification process used.
Figure 3
All 21 acetone sources were found to contain four of the common acetone impurities expected; diacetone alcohol (2-pentanone,4-hydroxy-4-methyl-), mesityl oxide (3-penten-2-one,4-methyl-), phorone (2,5-heptadien-4-one,2,6-dimethyl-) and phenol, but at varying amounts. Figure 4 illustrates a heatmap plot summarizing the results obtained from these four compounds for the 21 acetone sources using the A×C column set. In Figure 4, the solute zone volumes (absolute detector counts) are based upon the average of three replicate injections of the acetone source, which were analysed in a random order. The average solute volumes illustrated in Figure 4 are calculated based on a log10 scale, where the blue color represents the lowest signal intensity and the red colour represents the largest signal intensity of the compound. As illustrated in Figure 4 there appears to be a large variation in the amount of these four compounds for each acetone source. This indicates that acetone CIS injection process is not contributing appreciably to the presence of these compounds and that the components are likely residual from the manufacturing or distribution process. From Table 1, acetone sources A23, A24 and A25 are from the same distributor and the same lot (i.e., different bottles); acetone sources A9 and A10 are from the same distributor, but different lots (this is also true for acetone sources A5 and A6; however, A5 displayed a large isopropyl alcohol impurity of approximately 1:1 relative to acetone, chromatogram not shown); acetone sources A11, A27, A28 are commonly available hardware store brands. These 'source' trends are also observed in Figure 4, where these acetone sources tend to cluster closer to each other. The exception maybe the closer clustering of A9 and A20 to each other compared to the A9 and A10. Again this clustering is only using these four expected impurities, where the actual number of compound impurities is extremely large. As shown in Figure 2 the number, type and amount of impurities present between these two sources (A9 and A20) is very different.
Figure 4
Conclusion
For broad spectrum screening or trace component detection against complex background mixtures, impurities present within a solvent can significantly confound research efforts. In this work comprehensive gas chromatography–mass spectrometry is used to assess the impurities present in various commercial acetone brands. A large number and broad range of different chemical impurities were found, dependent upon the acetone source. These chemical impurities ranged in complexity from simple organic acids, alkanes and olefins, to a variety of substituted benzyl compounds (to highlight just a few). In general, dependent upon the analytical questions trying to be answered, a chemist should not become complacent and should be cautious of the impurities that may be present (and how they may change with different brands) and interfere with a trace component analysis. This is highlighted by the presence of the organic acids present in various acetone sources and the subsequent need for two different column sets to better capture and understand the range of impurities present. Moreover this work addresses only those compounds that are stable and volatile for the gas chromatographic system and does not address those components more labile or nonvolatile that may also be present in a solvent to alter or affect results.
Acknowledgements
Funding for this work was provided through contract AGRHSHQDC07X00451 to Pacific Northwest National Laboratory by the United States Department of Homeland Security, Science and Technology Directorate.
Jon Wahl is a senior research scientist in the Chemical and Biological Signature Sciences Group at Pacific Northwest National Laboratory. He has a PhD in Analytical Chemistry from Michigan State University and a BS in Chemistry from Michigan Technological University. Dr Wahl's principal research interests focus on interdisciplinary chemical and bioanalytical problems that require advanced high-resolution separation techniques. Multidimensional separations for ultratrace analysis methods for chemical warfare agents and related compounds, biological, explosives and other signature compounds of interest are a main focus.
Cinnamon Bolz received a BS in Biochemistry from Portland State University and a MS in Forensic Science from National University. She has experience in a variety of analytical techniques including separations, mass spectrometry and spectrophotometry in addition to writing standard operating procedures and quality control manuals, and applying statistical data analysis.
Karen Wahl is a staff scientist within the Chemical and Biological Signature Sciences Group at Pacific Northwest National Laboratory in Richland, Washington, USA. Dr Wahl has a PhD in Analytical Chemistry from Michigan State University and a BA in Chemistry from The College of Wooster. Dr Wahl's research interests include the development of MALDI mass spectrometry for microorganism analysis and the development of analytical methods for chemical and biological forensics.
References
1. V.J. Aristovich et al., Method for Purifying Acetone, US Patent 6340777.
2. K. Hunchak and I.H. (Mel) Suffet, J. Chromatogr., 392, 185–198 (1987).
3. A.L. Lafleur and N. Pangaro, Analytical Letters, 14(A19), 1613–1624 (1981).
4. H.W. Kreuzer et al., J. Forensic Sci., In Press.
5. P.J. Marriott and R.A Shellie, Trends Anal. Chem., 21, 573–583 (2002).
6. J. Beens and U.A.Th. Brinkman, Analyst , 130, 123–127 (2005).
7. M. Adahchour et al., Trends Anal. Chem., 25, 540–553 (2006).
8. M. Adahchour et al., Trends Anal. Chem., 25, 726–741 (2006).
9. M. Adahchour et al., Trends Anal. Chem., 25, 821–840 (2006).
Articles in this issue
over 16 years ago
Event Newsover 16 years ago
Method Development for Drug Impurity Profiling: Part 1over 16 years ago
Getting the Most Out of Calibration StandardsRelated to this article


