LCGC Advances In Gas Chromatography 11.23-11.24
|Articles|December 2, 2008

The Application Notebook

  • The Application Notebook-12-01-2008
  • Volume 0
  • Issue 0

maXis High Resolution LC–MS Makes the Most of Ultrafast LC Separations

An ultrafast gradient LC separation method was developed to separate a 5-component drug mixture in 30 seconds, with peak widths of 1 second. maXis mass accuracy at sub-ppm levels and true isotopic pattern of the spectra from the peaks lead to a confident elemental formula assignment for each drug compound with the SmartFormula algorithm.

Catherine Stacey, Sebastian Goetz and Carsten Baessmann, Bruker Daltonik GmbH, Bremen, Germany.

An ultrafast gradient LC separation method was developed to separate a 5-component drug mixture in 30 seconds, with peak widths of 1 second. maXis mass accuracy at sub-ppm levels and true isotopic pattern of the spectra from the peaks lead to a confident elemental formula assignment for each drug compound with the SmartFormula algorithm.

Introduction

Modern discovery applications demand definitive MS results on ever more complex samples. maXis is the only mass spectrometer able to deliver the maximum MS performance specification at the very highest speeds made available by modern ultrafast liquid chromatography. Here, a complete gradient LC separation of typical drug molecules is achieved in 30 seconds, with each component detected with sub-ppm mass accuracy.

Figure 1

Experimental

A drug mixture (norharmane, sulphamethazine, sulphamethoxazole, oxybutynin and terfenadine) was separated by an ultrafast separation LC (U-HPLC) using a flow-rate of 0.5 mL/min and a gradient from 4% acetonitrile to 95% acetonitrile in 0.5 mins. The maXis electrospray LC–MS dataset was acquired with an acquisition rate of 10 Hz. The dataset was automatically recalibrated in the Compass DataAnalysis 4.0 processing software with several masses of salt clusters resulting from a loop injection of lithium formate solution (10 mM in water/isopropanol) at the end of the separation.

Figure 2

Results

The separation of the 5 drug mix gave baseline resolution of each compound within 30 seconds (Figure 1). For each peak more than 15 spectral measurements were made across each peak, to give reproducible peak areas suitable for quantitative analysis (Figure 2). In the mass spectrum, the mass accuracy for each peak of the isotopes is below 1 ppm, with a resolution of 43700 — acquired at an acquisition rate of 10 Hz (Figure 3). SmartFormula analysis is automated in the Compass 1.3 data processing software. SmartFormula automatically determines the possible elemental compositions of each compound LC peak, using a mass accuracy of < 2 ppm error with allowed chemical rules and elements (CnHnNnOnSn). Formulae are ranked according to the closest match factor between the measured isotopic pattern and the theoretical pattern for a given formula — the Sigma value, which should be close to zero. For these drug compounds, a unique SmartFormula result was produced within the allowed mass accuracy and Sigma values (Figure 4). Peak 4, Oxybutynin (Figure 5) was unequivocally confirmed in this experiment.

Figure 3

Conclusion

The maXis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern ultrafast chromatography. Sub-ppm mass accuracy is shown for small molecule drugs, together with a resolution in excess of 40000. Thus the instrument allows extraction of full quantitative information from high speed chromatography. Supported by SmartFormula, the maXis provides unequivocal molecular formula determination capabilities. For compounds with higher molecular weight (up to 1000 Da) SmartFormula 3D is using MS–MS fragment data for certainty in small molecule identifications.1

Figure 4

For research use only. Not for use in diagnostic procedures.

Reference

1. Bruker Daltonics Technical Note-23, "Certainty in Small Molecule Identification by Applying SmartFormula 3D on a UHR-TOF Mass Spectrometer."

Bruker Daltonik GmbH

Fahrenheitstr. 4, 28359 Bremen, Germany

tel. +49 421 2205 0 fax +49 421 2205 104

Internal server error