Molecular Weight Determination of Unfractionated Heparin
Heparin is a polydisperse, heterogeneous polysaccharide derived from animal tissue. Heparin has been used as an anticoagulant for over 60 years and one of the fundamental parameters for characterizing unfractionated heparin (UFH) is its molecular weight (MW) and MW distribution. The molecular weight of heparin ranges from 3 to 50 kDa, but, typically falls within the 10 to 20 kDa range for unfractionated heparins used in medical applications. In light of heparin contamination issues in 2008, USP has proposed to include heparin molecular weight determination for the stage 3 heparin monograph revisions.
Heparin is a polydisperse, heterogeneous polysaccharide derived from animal tissue. Heparin has been used as an anticoagulant for over 60 years and one of the fundamental parameters for characterizing unfractionated heparin (UFH) is its molecular weight (MW) and MW distribution. The molecular weight of heparin ranges from 3 to 50 kDa, but, typically falls within the 10 to 20 kDa range for unfractionated heparins used in medical applications. In light of heparin contamination issues in 2008, USP has proposed to include heparin molecular weight determination for the stage 3 heparin monograph revisions.
Many high-performance size-exclusion chromatography (SEC) systems have been developed and represent the most commonly used method in the MW determination of UFH. At Scientific Protein Laboratories (SPL) we have developed an SEC system coupled with multi-angle light scattering (MALS) for the determination of absolute MW and MW distribution of UFH and heparin derivatives. MALS eliminates the need for well characterized molecular weight standards derived from heparin, and our procedure represents an improvement over the column calibration techniques used in other relative MW determination methods.
For this application note, this method uses an Agilent LCâ1200 series HPLC, 0.1 M ammonium acetate mobile phase, Shodex SB-804HQ analytical column (8.0 × 300 mm) with Shodex SB-G guard column, miniDAWN TREOS light scattering detector, and Optilab rEX refractive index detector. In addition to UFH, the method is capable of analyzing dermatan sulphate and over-sulphated chondroitin sulphate (OSCS) and can distinguish among the three (Figures 1 and 2). The specific refractive index increment (dn/dc) and second virial coefficient (A2) measurements can also be determined using the Wyatt Optilab and Wyatt miniDAWN TREOS detectors, respectively.
This note was graciously submitted by Lin Rao and John Beirne, Scientific Protein Laboratories, LLC.
Wyatt Technology Corporation
6300 Hollister Avenue, Santa Barbara, California 93117, USA
Tel: +1 (805) 681 9009 fax: +1 (805) 681 0123
E-mail:
Articles in this issue
over 11 years ago
Chromate in Toys, Leather, and Potable Waterover 11 years ago
Analysis of Organochlorine Pesticides (OCPs) on SGE BP5MSover 11 years ago
Fingerprinting Analysis of Different Types of Beerover 11 years ago
Fast and Simple Determination of Free Amino Acids in Milkover 11 years ago
Membrane Proteinsover 11 years ago
Glycosylation Analysis with TSKgel Amide-80 2 µm UHPLC ColumnsRelated to this article