LCGC Advances In Gas Chromatography 11.23-11.24
|Articles|November 20, 2015

LCGC Europe

  • LCGC Europe-12-01-2015
  • Volume 28
  • Issue 12

Practical Applications of Asymmetrical Flow Field-Flow Fractionation (AF4): A Review

Characterization of macromolecules and colloids is an area of considerable interest. Asymmetrical flow field-flow fractionation (AF4) has become a well-established method, but many potential users possess limited knowledge of its capabilities, and that it can provide additional information or serve as validation to the traditional analytical techniques. This review article highlights several practical applications in which AF4 should be given special consideration, and discusses benefits and drawbacks of the different methods.

Characterization of macromolecules and colloids is an area of considerable interest. Asymmetrical flow field-flow fractionation (AF4) has become a well-established method, but many potential users possess limited knowledge of its capabilities, and that it can provide additional information or serve as validation to the traditional analytical techniques. This review article highlights several practical applications in which AF4 should be given special consideration, and discusses benefits and drawbacks of the different methods.

  It is almost 50 years since the late Professor Giddings (1) invented field-flow fractionation – a family of separation techniques. A number of different sub-techniques exist; centrifugal, thermal, and electrical to name a few, each with its own appropriate application area. However, the most widely spread and commonly used sub-technique is asymmetrical flow field-flow fractionation (AF4). Since the first reports of AF4 in the 1980s (2) the technique has evolved to become a well-established method for size separation of macromolecules and particles in academia. The technique has also found its way into the industry, largely because of the availability of commercial systems in the latter half of the 1990s.    However, many potential users have still not heard about AF4 or refrain from using it because they have little experience with the technique. With the advent of improved instrumentation and an ever-increasing number of articles to help to select separation parameters, the relevance of AF4 is being re-evaluated. In this review article we highlight applications where AF4 is especially well suited in comparison to other common techniques.  

About AF4 (Facts)

  More detailed descriptions on how asymmetrical flow field-flow fractionation (AF4) works can be found in other

LCGC

articles (3,4) and in several books on the subject (5,6) In short, AF4 is a size separation technique for macromolecules, aggregates, particles and other colloids that are dispersed in a solvent. The separation takes place in an open channel (no stationary phase) in which one wall consists of a semi-permeable ultra-filtration membrane. The open channel give low surface area, low shear forces, and the ability to separate very large-sized components. In combination with the on-line coupling of light scattering detection AF4 provides information on molar mass and size on the separated fractions directly without calibration standards.  

Detection of Protein Aggregates 

  Studies of protein aggregation are common in both academia and industry. Protein aggregation can be of fundamental interest to understand the characteristics of a protein, but is also important to assess during product development and optimization, or to study the stability of the finished formulation. Size-exclusion chromatography (SEC) is a well-proven and excellent method for size separation of macromolecules, and is often used for size characterization of proteins including aggregation studies. However, it has been reported that SEC has limitations for the protein aggregates, indicating that the method is not always well suited for the separation and detection of higher aggregates (7,8,9), which can result in low recovery (10).   An example of differences that can be noticed in the results from SEC and AF4 analyses of an aggregated protein is presented in Figure 1. The sample is a solution of IgG (molecular weight ~150 kDa) that is well-known to aggregate at elevated temperatures. For this illustration the solution was heated to ~65 ºC for 5 min to promote the formation of aggregates. Elution of the protein monomer and dimer from AF4 was detected on both the differential refractive index (dRI) and multi-angle light scattering (MALS) detectors. The light scattering detector, which is very sensitive to high molar mass compounds, also picked up the elution of much higher aggregates (not detected in the IgG solution before heat-stress). The amount of these aggregates is low compared to the monomer and dimer, representing less than 10% of the total protein. In contrast to the AF4 results, no higher aggregates were detected in the analysis by SEC. This is presumably a result of either shear degradation of the higher aggregates or that the aggregates do not elute from the column.    The SEC column had a nominal pore-size of 1000Å, a nominal protein molecular weight separation range of 50–7500 kDa, and was operated at the relatively low flow rate of 0.5 mL/min (7.8 mm i.d. column). Hence, the protein and potential aggregates are expected to be well within column specifications. Nevertheless, the higher protein aggregates were not eluted from the column. Optimization of elution conditions such as the mobile phase ionic strength and other additives (10 mM sodium nitrate in water was used) and testing of a variety of columns with different pore-sizes and chemistries could improve the situation. However, it is not certain that one would go through such an optimization procedure unless there is an indication that the current SEC method may fail to detect the aggregates. Great care should therefore be taken when estimating protein aggregation based on data from one method alone; the use of a complementary technique can be valuable for validation reasons   

Size Distribution of Nanoparticles 

  Nanoparticles are becoming more and more closely studied. One property of high interest is the particle size, which is commonly analyzed with dynamic light scattering (DLS). This technique measures the fluctuations in scattered light from a sample (such as a particle dispersion) over time and provides information on the hydrodynamic diameter. DLS has several benefits such as compatibility with most solvents and is a relatively fast technique with low workload for the user. However, in some cases, it is difficult to obtain reliable size distribution data (11) on complex samples such as particles with multi-modal (12) or broad size distributions (13), or when measuring small particles at low concentrations. This can be explained by the fact that an inherent property of DLS is that the measurements are performed on unfractionated samples, making the data processing and evaluation more complicated for samples with broad size distributions. For small particles, the limitation lies in the relatively low scattering intensity (the intensity of scattered light increases with particle size by the power of 6), which becomes especially challenging if the concentration is low.    The following example is from a recent project in which we participated where DLS was used to characterize the size of nanoparticles. Our DLS measurement gave an average hydrodynamic size of 73 nm and a polydispersity index of 0.25, which was in excellent agreement with the data given by the supplier; 72 nm and 0.25, respectively (also determined by DLS). Looking at the intensity graph from the DLS (Figure 2) it is obvious that the size distribution is broad (ranging from 10–400 nm) but there was otherwise no indication of anything remarkable. Possibly there is an indication of a shoulder on the size distribution curve at approximately 20–30 nm.   In contrast, AF4 analysis of the same sample gave a more detailed picture of the size distribution (Figure 3). The use of MALS-dRI detection gave molar mass and r.m.s. radius (a measure of [a molecules or particles] size weighted by the mass distribution around its centre of mass) information directly of the eluted fractions (independent of the separation and no standards necessary). The analysis detected three size populations on the refractive index and light scattering detectors. The first population (I) eluted from 2–5 min, the second (II) from 5–10 min, and the last (III) was visible mainly in the MALS-signal from 11–18 min. The size populations were incompletely resolved, partly as a result of band broadening, but a major factor is likely overlapping size distributions. This particle was produced in a two-step process, with a solid core that was coated. The core had a size of 5–10 nm as determined with transmission electron microscopy (TEM) prior to coating (likely population I in Figure 3). The second size population (II) probably corresponds to successfully coated particles and the third size population (III) to aggregated particles.   DLS is a rapid tool, excellent for fast screenings of well defined, low dispersity samples. However, this data illustrates the difficulties involved in detecting different size populations in a complex sample using DLS and is one example when it could be suitable to characterize the sample with AF4. The use of a separation technique also offers the potential for an additional dimension because the size fractions can be collected and analyzed by other techniques to further characterize the sample.  

Molar Mass and Size Determination of Large Polysaccharides 

  Polysaccharides are a class of compounds that are used in a wide variety of applications. Some of the polysaccharides, such as starch and cellulose, can be very large and are often used for rheological modification of products. However, the large sizes can also result in challenging size distribution analysis (14,15). Often only average properties are reported, or indirect measurements of the size such as viscosity. However, with AF4 it is possible to analyze very large polysaccharides (16,17,18,19) as well as supramolecular aggregates (20,21,22). Size analysis can be important because the molar mass distribution can be affected during functionalization or by processing conditions (23,24,25,26,27,28).    The analysis of amylopectin (one of the main constituents of starch) by AF4 is presented in Figure 4. Amylopectin is a branched polysaccharide of very large size (hydrodynamic diameter of several hundreds of nm) and molar mass (average molar mass commonly in excess of 10

7

g/mol). Molar mass ranges from approximately ten million up to several hundred millions g/mol and r.m.s. radius ranges up to ~250 nm corresponding to a hydrodynamic diameter of 400–500 nm, which emphasizes the large size range over which AF4 can operate.   In this case AF4 was used to monitor dissolution conditions. Amylopectin is not easily dissolved in water, and the dissolution conditions can have a pronounced effect on the properties of the final amylopectin solution. In this example the dissolution temperature and its impact on the size distribution was investigated and compared to dissolution in dimethyl sulphoxide (DMSO) (Figure 5). The AF4 analysis clearly revealed the impact of high temperature dissolution conditions on the amylopectin (degradation) and provided guidance on how to choose the amylopectin processing conditions for the particular application.   Using SEC for large compounds as in the example above would definitely be challenging. The size of the amylopectin would likely be excluded from the pores of the stationary phase. Furthermore, even if the pores are sufficiently large enough to allow the amylopectin to enter, the flow rate has to be kept very low to allow the sample time to diffuse in and out of the pores. In addition, shear degradation of amylopectin readily occurs as has been shown by Cave

et al.

(29).  

Size Analysis of Vesicles/Liposomes

  Liposomes as carriers for pharmaceutically active compounds have attracted both academic and commercial interest for several decades. One parameter that really matters in drug applications of liposomes is the size distribution because it has a considerable impact on the clearance of the liposomes from the blood stream (30,31,32).   AF4 is a technique very well suited for liposome characterization (33,34,35,36,37,38). The open channel reduces the risk of shear-induced changes to the sample and has a very low surface area (typically in the order of 10

-3

–10

-2

m

2

) compared to packed columns (at least 1–10 m2/g stationary phase). An example of an AF4 analysis of liposomes is presented in Figure 6. The liposome size distribution was investigated after different processing steps for different liposome preparation procedures.    The AF4 separation demonstrated good recovery (98%) and was able to separate the relatively disperse systems, with liposome distributions spanning from ~25 to ~500 nm depending on production protocol. The use of a separation method also allowed different liposome size fractions to be collected after the AF4, and analyzed for their compositions by, for example, liquid chromatography–mass spectrometry (LC–MS).   The most commonly used techniques to analyze size and size distribution of liposomes are DLS and SEC (39). DLS suffers from the same limitations as was mentioned in the nanoparticle example above (difficulty to obtain accurate size distribution data for broad or complex size distribution), and this has also been noted for liposomes (35). SEC differs from DLS in that it give a physical separation of the differently sized liposomes, enabling direct measurement of the size distribution (40,41,42,43). However, lipid adsorption to the SEC column is often reported (41,43). To counter this, it is recommended to (41) the column by several injections of lipid before the actual separation can be performed. Obviously, saturating the stationary phase prior to separation might affect column performance and retention properties. Therefore, using AF4 in such cases might be a better option.  

Separation of Branched Polymers

  Branched polymers are a widely used class of macromolecules. The amount of branching as well as the molar mass is of considerable interest since it can have a profound affect on the properties of the polymer in the application. Both parameters can be investigated by SEC when connected to viscosity and light scattering detectors. However, it is not uncommon to see reports with anomalies in the elution behaviour when analyzing branched polymers (44,45,46,47,48,49,50,51), making the evaluation challenging or even impossible. Several investigations point at the difficulties of separating branched high molar mass polymers. One suggested explanation has been anchoring of the branches in the SEC stationary phase pores (47).   For polymers of this complex character, AF4 is an attractive alternative. One illustrative example of using AF4 for separation of branched polymers can be seen in Figure 7 (including a comparison with SEC). The absence of a packing material in AF4 significantly reduces the risk for shear-induced degradation or branches interacting with a stationary phase.   

When Not to Use AFFF – and What Not to Expect

  Compounds smaller in size than the cut-off of the membrane used are not possible to analyze. The most commonly used membranes have a nominal molecular weight cut-off of 5 kDa, or 10 kDa, which gives an indication of the practical lower limit of AF4. SEC can be expected to exhibit better efficiency or fractionating power than field-flow fractionation (FFF) for low-molecular-weight compounds (~ 10

4

g/mol range and less) (52,53,54).   If the amount of sample is very limited, the use of AF4 might not be possible. The amount injected (in terms of mass) is comparable to what is used in SEC, usually 10–100 µg depending on detection. However, one benefit of AF4 is that if the sample concentration is low, larger volumes can be analyzed (thus reaching the required detection) without affecting resolution thanks to the initial focusing step when the analytes are relaxed in the channel.   The length of an AF4 separation varies depending on sample complexity. In general, total AF4 analysis time (injection, focusing, and elution) is in the 15–60 min range. Add to this the time for data evaluation, which can be very fast for a simple yes/no answer or take considerably more time if data from multiple detectors are to be evaluated and sample characteristics in terms of conformation/branching/density are desired. Therefore, answers are not obtained as fast as with DLS; however, AF4 offers considerably shorter run times compared to analytical ultracentrifugation.   While AF4 does separate according to size, which in turn is related to mass, do not expect mass spectrometry-like data with isotope resolution. The resolution is simply not of that magnitude.   

Conclusion

  We have offered some examples where AF4 should be considered in comparison to other commonly used techniques, either because it can provide additional or complementary information or can serve as validation of other analytical techniques. Speaking in general terms, AF4 should be considered when the analytical task involves:   - Aggregation-prone samples or when investigating aggregation and aggregate properties; - Broad or complex size distributions; - Large macromolecules; - Shear sensitive samples; - Compounds that are difficult to analyze because of interactions with stationary phases.  

References

  1. J.C. Giddings, Separation Sci. 1, 123–125 (1966). 2. K.-G. Wahlund and J.C. Giddings, Anal. Chem. 59, 1332–1339 (1987). 3. S. Podzimek et al., LCGC North America 27(1), 62–69 (2009).  4. R.N. Qureshi and W.Th. Kok, LCGC Europe 23(1), 18–25 (2010).  5. K.-G. Wahlund, Asymmetrical flow field-flow fractionation in Field-flow fractionation handbook, M.E. Schimpf, K.D. Caldwell, and J. C. Giddings, Eds. (Wiley, 2000).  6. K.-G. Wahlund and L. Nilsson, Flow FFF – basics and key applications in Field Flow fractionation in biopolymer analysis, K. Williams and C. Caldwell, Eds. (Springer, 2012).  7. J.S. Philo, Curr. Pharma. Biotech. 10, 359–372 (2009). 8. J.P. Gabrielson et al., J Pharm Sci 96, 268–279 (2007). 9. S. Cao et al., Curr Pharm Biotechnol. 10(4), 382–390 (2009). 10. A. Hawe et al., Eur J Pharm Sci. 38, 79–87 (2009). 11. Kato et al., Nanomaterials 2, 15–30 (2012). 12. C. Franzen et al., AAPS Pharm Sci Tech, 4, 62–70 (2003).  13. M. Baalousha and J.R. Lead, Environ Sci Tech 46, 6134–6142 (2012). 14. W. Yokoyama et al., Cereal. Chem. 75(4), 530–535 (1998). 15. R.A. Cave et al., Biomacromolecules 10(8), 2245–2253 (2009). 16. L. Nilsson, Food Hydrocolloids 30, 1–11 (2013). 17. L. Nilsson, Starch and other polysaccharides in Field Flow fractionation in biopolymer analysis, K. Williams and C. Caldwell, Eds. (Springer, 2011).  18. P. Roger et al., J. Chrom. A 917, 179–185 (2001). 19. K.G. Wahlund et al., Anal. Bioanal. Chem. 399(4), 1455–1465 (2011). 20 A. Håkansson et al., Carb. Polymer 87, 518–523 (2012).  21. M. Andersson et al., Anal. Chem. 73(20), 4852–61 (2001). 22. M. Andersson et al., Biomacromolecules 5(1), 97–105 (2004).  23. J. R. Runyon et al., Analytical and Bioanalytical Chemistry 406, 1597–1605 (2014).  24. S. Lee et al., J. Chrom. A 1011, 111–123 (2003).  25. G. Modig et al., Food Hydrocoll. 20, 1087–1095 (2006).  26. L. Nilsson et al., Biomacromolecules 7(9), 2671–2679 (2006).  27. M. Ulmius et al., Food Hydrocolloids 26(1), 175–180 (2012). 28. D. Perez-Rea et al., Anal Bioanal Chem 407(15), 4315–26 (2015). 29. Cave et al., Biomacromolecules 10(8), 2245–53 (2009). 30.R.L. Juliano and D. Stamp, Biochem. Biophys. Res. Commun. 63, 651–658 (1975). 31. R.M. Abra and C.A. Hunt, Biochim Biophys Acta 666, 493–503 (1981). 32. Liu et al., Biochim Biophys Acta 1104(1), 95–101 (1992).  33. M.H. Moon and J.C. Giddings, J. Pharm. Biomed. Anal. 11, 911 (1993). 34. Moon et al., Journal of Chromatography A 813, 91–100 (1998). 35. Hupfeld et al., J. Nanosci. Nanotech. 6, 1–7 (2006). 36. G. Yohannes, Dissertation thesis, University of Helsinki, Finland (2007). 37. J. Kuntsche, C. Decker, and A. Fahr, J Sep Sci. 35(15), 1993–2001 (2012).  38. Wyatt Technology Corporation, LCGC The Application Notebook, (2011).  39. Laouini et al., J. Coll. Sci. Biotech. 1, 147–168 (2012). 40. S. Lesieur, C. Grabielle-Madelmont, M.T. Paternostre, and M. Ollivon, Chem Phys Lipids 64, 57–82 (1993). 41. Lundahl et al., J Chrom B, 720, 103–120 (1999).  42. Grabielle-Madelmont et al., J Biochem Biophys Methods 56(1–3), 189–217 (2003).  43. Ruysschaert et al., BMC Biotech. doi:10.1186/1472-6750-5-11 (2005). 44. C. Johann and P. Kilz, J Appl Polym Sci Appl Polym Symp. 48, 111–122 (1991).  45. D.J. Frater et al., J Polym Sci. B 35(1), 141–151 (1997).  46. M. Gerle Fischer et al., Macromolecules 32, 2629–2637 (1999). 47. S. Podzimek et al., Journal of Applied Polymer Science 81(7), 1588–1594 (2001).  48. E. Mes et al., Journal of Chromatography A 1154(1–2), 319–330 (2007).  49. T. Otte et al., Journal of Chromatography A 1218(27), 4240–4248 (2011).  50. Otte, T et al., Journal of Chromatography A 1218(27), 4257–67 (2011).  51. A. Makan et al., Macromolecules 45(12), 5247–5259 (2012). 52. J. J. Kirkland et al., Analytical Chemistry 52, 1944–54 (1980).  53. M.E. Schimpf, Resolution and fractionating power in Field-flow fractionation handbook, M.E. Schimpf, K.D. Caldwell, and J.C. Giddings, Eds. (Wiley, 2000).  54. K.D. Caldwell, Polymer analysis by field-flow fractionation in Modern methods of polymer characterization, H.G. Barth and J.W. Mays, Eds. (Wiley, 1991).  

Mats Leeman

is Senior scientist at SOLVE Research and Consultancy AB. Mats holds a PhD in Technical Analytical Chemistry from Lund University where he studied under the supervision of Professor K.-G. Wahlund. He has been using AF4 for more than 15 years for various applications within the pharmaceutical, specialty chemicals, and biotech industries.  

Matilda Ulmius Storm

is the CEO of SOLVE Research and Consultancy AB, a contract analysis company focusing on material characterization using AF4 among other techniques. The company is based in Lund, Sweden, and Matilda holds a PhD in biomedical nutrition/pure and applied biochemistry from Lund University where she used AF4 to understand the properties and behaviour of beta-glucans. She has experience in using AF4 technology for various applications involving polysaccharides, proteins, and nanoparticles.  

Lars Nilsson

is currently Associate Professor in Food Technology at Lund University. He earned his PhD at Lund University in 2007 working on the adsorption of macromolecules during emulsification. After a brief time in the biotech industry, working on formulation and stability of foams, he returned full-time to Lund University in 2009. In 2010 Lars was awarded the Akzo Nobel Nordic Price in Surface and Colloid Chemistry for his work on macromolecular emulsifiers. Current research interests are field-flow fractionation techniques for macromolecules and colloidal dispersions, and structure/functionality relationships in macromolecules and their aggregates, emulsions, and foams as well as brewing technology.


Related to this article