High Speed Seperation of More Than 16 Polycyclic Aromatic Hydrocarbons

Article

LCGC Asia Pacific

LCGC Asia PacificLCGC Asia Pacific-09-01-2009
Volume 12
Issue 3

Knauer Application Note

S. Marten and A. Knöfel, Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH, Berlin, Germany

Introduction

Polyaromatic hydrocarbons (PAH) are particularly relevant in the analysis of environmental pollution because of their ubiquity, toxicity and persistence. Consequently, the PAHs have become the most intensively studied pollutants in environmental analysis.1,2 Standard and official methods for their analysis are available in guidelines for air, water, solid waste and food analysis.3 To attain the required selectivity of HPLC separations stationary phases specifically designed for PAH analysis are required.

Using standard HPLC for PAH analysis, injection-to-injection cycle times can often exceed 30 minutes. Under optimized conditions, gradient separations of the US EPA recommended standard test mixture of 16 PAH can be performed in about 15 minutes. The aim of this study was to investigate the resolution limit and analysis time for 16 PAH of the EPA standard plus 3 additional related aromatic compounds (1-methylnapthalene, triphenylene and p-terphenyl) using a PLATINblue system with a short BlueOrchid PAH column.

Experimental

The PAH analysis was developed on BlueOrchid PAH 50 × 2 mm column and a PLATINblue UHPLC high pressure gradient system. Please refer to Figure 1 for details.

Figure 1: Separation of 19 PAHs, overlay of five runs. Column: BlueOrchid PAH, 50 × 2 mm.

Results

After method optimization for the separation of the standard PAH mixture we spiked the mixture with 3 additional PAHs. To achieve a separation of these 19 PAHs in less than 2 min the method parameters were slightly modified (see Figure 1). In spite of the fast gradient applied (composition change rate of 2% per second), the reproducibility of the method was good with a maximum deviation of ΔtR = 0.012 min. In summary, we would recommend a flow-rate of 1 mL/min for an analysis time of 2.5 min, plus an equilibration time of 0.4 min run at a higher flow-rate of 1.5 mL/min. Side-effect: compared with the standard HPLC method the method above accounted for an eluent savings of 80% per analysis.

Conclusion

By using a shorter column with a smaller inner diameter and a KNAUER PLATINblue UHPLC system, complex mixtures such as the 19 PAHs can be analysed with excellent reproducibility and cycle times of less than 3 minutes. Optimizing the speed and resolution of routine analyses can not only save time but also dramatically decrease eluent costs, particularly important for analyses using acetonitrile.

References

1. M.N. Kayali-Sayadi et al., Fresenius J Anal Chem., 368(7), 697–701 (2000).

2. T. Wenzl et al., Trends in Analytical Chemistry, 25(7), 716–725 (2006).

3. AOAC 973.30; Deutsche DIN TVO; UK ISBN 0 11 752032 2; US EPS Methods TO-13, 550 550.1, 610, 8310 8330.

Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH

Hegauer Weg 38, 14163 Berlin, Germany

tel. +49 30 809727 0 fax +49 30 8015010

E-mail: info@knauer.net Website: www.knauer.net

Related Videos
Robert Kennedy
John McLean | Image Credit: © Aaron Acevedo