LCGC Advances In Gas Chromatography 11.23-11.24
|Articles|December 1, 2010

LCGC Supplements

  • Special Issues-12-01-2010
  • Volume 0
  • Issue 0

Isolation and Purification of Ranitidine Using Xselect Columns and Focused Gradients

Isolation and purification of compounds and their impurities/degradants is an important part of the drug discovery and development process.

Kenneth J. Fountain, Jo-Ann M. Jablonski, and Damian Morrison, Waters Corporation

Isolation and purification of compounds and their impurities/degradants is an important part of the drug discovery and development process. It is critical that a sufficient mass of each compound be isolated and purified for subsequent analysis and testing in the least amount of time to ensure timely identification and characterization. However, developing a chromatographic separation at the preparative scale can be time-consuming and costly due to solvent use and disposal, run time, and potential loss of crude material during the gradient optimization process. As an alternative, analytical columns (4.6 mm inner diameter) can be used to quickly develop analytical separations using mobile phases that are compatible with preparative chromatography. A focused gradient developed at the analytical scale can be easily transferred to the preparative scale using the same stationary phase chemistry. Focused gradients improve the resolution between closely eluting peaks without increasing the run time, leading to higher purity and increased yield of the target product. In this application, ranitidine, the active ingredient in Zantac, is isolated and purified at the preparative scale based on the analytical separation results. The collected fractions are then analysed with the focused gradient method to rapidly confirm purity. Selectivity of the separation at both the analytical and preparative scales is maintained by using the same chemistry packed into two different column dimensions. The XSelect columns used in this application are stable at both high and low pH, which allows them to be used with a wide range of separation conditions and mobile phases compatible with preparative chromatography.

Experimental

System: AutoPurification system with 2996 PDA and AutoPurification flow cell

Analytical column: XSelect CSH C18, 5 μm, 4.6 × 100 mm

Prep column: XSelect CSH Prep C18 OBD, 5 μm, 19 × 100 mm

Mobile phase A: 0.1% NH4OH in H2O

Mobile phase B: 0.1% NH4OH in CH3OH

Temperature: 30 °C

Detection: UV 240 nm

Sample concentration: 72.7 mg/mL

Sample diluent: DMSO

Injection volume: 1 μL for crude and 20 μL for purified fractions (analytical); 119 μL (prep)

Results and Discussion

A generic screening gradient was used to analyse ranitidine, the active ingredient in Zantac, which reduces the acid levels in the stomach and is used to treat ulcers and gastroesophageal reflux disease (GERD). The resulting separation is shown in Figure 1 (top). Since ranitidine is a base with one of its pKa values at 8.2, it has better retention under high pH mobile phase conditions. Ammonium hydroxide was chosen as the mobile phase modifier because it can be removed more easily from purified fractions at the preparative scale when compared to bicarbonate or acetate buffers.

Figure 1: Analysis of ranitidine using an analytical screening gradient (top), an analytical focused gradient (middle) and a preparative focused gradient (bottom). The screening gradient was from 5 to 95% B in 12 min. The focused gradient had an initial hold at 5% B for 1 min, to 35.1% B in 0.5 min, then to 43.1% B in 5.36 min, followed by a rinse at 95% B for 1 min. Analytical flow rate was 1.46 mL/min. Preparative flow rate was 25 mL/min.

Based on the chromatography from the screening gradient, a focused gradient was developed on the analytical column to ensure all impurities were more sufficiently resolved from ranitidine (1). In order to develop a focused gradient, several parameters must be calculated. The calculations shown below describe the step-by-step process needed to formulate a focused gradient from the results obtained in the screening gradient:

Corrected retention time (RT) of ranitidine = Retention time – time to detector* – gradient hold

=6.30 min – [(0.93 mL + 1.395 mL)/1.46 mL/min] – 0 min

=4.71 min

*time to detector in min = (system volume in mL + column volume in mL)/flow rate in mL/min

Elution % of ranitidine = (Corrected RT/gradient time) × (% change over the gradient) + initial gradient%

= (4.71 min/12 min)(95% – 5%) + 5%

=40.1% (%MeOH for ranitidine elution)

Gradient slope expressed as % change per column volume = % change/[(flow rate in mL × gradient time in min)/column volume in mL]

=7.17% per column volume

Slope for focused gradient expressed as % change per column volume = scouting gradient slope/5

=7.17%/5

=1.43% per column volume

The focused gradient should start at 5% lower organic than the elution % and end 3% above the elution % for ranitidine. Therefore, the focused gradient for ranitidine should be from 35.1–43.1% methanol.

Focused gradient segment time = (% change in slope × column volume)/(focused gradient slope × flow rate)

=(8% × 1.395 mL)/(1.43% × 1.46 mL/min)

=5.36 min

The focused gradient for ranitidine on the analytical column is from 35.1–43.1% methanol in 5.36 min. Figure 1 (middle) shows the resulting separation for the crude ranitidine mixture using this gradient. Note that the ranitidine peak is better resolved from the surrounding peaks than in the scouting gradient. The focused gradient separation on the analytical column was then scaled to the preparative scale. The gradient was adjusted for gradient delay volume and gradient times were adjusted so that the number of column volumes for the duration of the gradient remained constant (1). The preparative separation can be seen in Figure 1 (bottom). The selectivity on the prep column is identical to the analytical column (middle), thus facilitating easier peak detection and fraction collection.

A total of 8.65 mg of the crude ranitidine mixture was injected onto the prep column and collected in two fractions. The fractions were then analysed on the analytical column using the focused gradient shown in Figure 1 (middle). Figure 2 shows the overlay of the two purified fractions and the crude starting material. The resulting purity of both fractions was greater than 99% (middle and bottom chromatograms). These fractions can now be evaporated and used in the next step of the drug development process.

Figure 2: Analysis of crude (top) and purified (middle and bottom) ranitidine using an analytical focused gradient. The focused gradient had an initial hold at 5% B for 1 min, to 35.1% B in 0.5 min, then to 43.1% B in 5.36 min, followed by a rinse at 95% B for 1 min. Flow rate was 1.46 mL/min.

Conclusions

An analytical method using a focused gradient was developed on an XSelect CSH C18 column for the analysis of ranitidine and its associated impurities/degradants. This gradient was then directly transferred to a preparative XSelect CSH Prep C18 OBD column for purification of ranitidine. The resulting fractions were analysed using the analytical method to confirm purity. Directly transferring chromatographic separations between the analytical and preparative column dimensions is accomplished by using the same stationary phase in a different column format. The use of focused gradients allows for high resolution separations without increasing run time. Information about focused gradient parameters can be directly obtained from the analytical screening gradient.

References

(1) J.M. Jablonski, T.E. Wheat, and D.M. Diehl, Developing focused gradients for isolation and purification, Waters application note, 720002955EN (2009).

©2010 Waters Corporation. Waters, XSelect, CSH Technology, OBD Technology, and The Science of What's Possible are trademarks of Waters Corporation. Zantac is a trademark of Glaxo Wellcome plc.

Waters Corporation

34 Maple Street, Milford, MA 01757

tel. (508) 478-2000; Fax (508) 478 -990

Website: www.waters.com

Articles in this issue

almost 16 years ago

Phenomenex

almost 16 years ago

MicroLiter Analytical Supplies, Inc.

almost 16 years ago

Pickering Laboratories, Inc.

almost 16 years ago

Bruker Daltonics

almost 16 years ago

Hitachi High Technologies America, Inc.

almost 16 years ago

Polymicro Technologies

almost 16 years ago

Analytical Sales & Services, Inc.

almost 16 years ago

J.G. Finneran Associates, Inc.

almost 16 years ago

KNAUER

Related to this article