
- The Application Notebook-02-01-2010
- Volume 0
- Issue 0
Analysis of Impurities in Fine Chemical Octyl-dimethyl-4- aminobenzoate Using the Agilent 1290 Infinity LC and ZORBAX RRHT and RRHD 1.8 µm Columns
The combined benefits are demonstrated by a separation of impurities found in a sample of octyl-dimethyl-4-aminobenzoate.
Michael Woodman, Agilent Technologies, Inc.
The combined benefits are demonstrated by a separation of impurities found in a sample of octyl-dimethyl-4-aminobenzoate. The high pressure capability of the system allows the use of methanol, as well as acetonitrile, to explore the selectivity of the two solvents. At 1.5 mL/min, using a simple 2-min gradient and a 3.0 mm X 50 mm 1.8 µm column, the analysis time is only 3 min. The separation of the main components is shown in Figure 1.
Figure 1
The speed, resolution and flexibility of the system are further demonstrated by a separation of the sample using methanol or acetonitrile with low solvent consumption 2.1 mm id, 1.8 µm columns. The flow rate and gradient conditions are optimized for each solvent, to produce a gradient separation with maximum pressure of approximately 850 bar, a conservative setting for the 1200-bar capability of the Agilent 1290 Infinity LC. The separation of the main components, with the two organic solvents, is shown in Figure 2a (acetonitrile, top panel) and 2b (methanol, lower panel), where the chromatograms are zoomed to the region of peaks shown from approximately 1.2–2.5 min in Figure 1.
Figure 2
Configuration
- G4220A 1290 Infinity Binary Pump with Integrated Vacuum
Degasser
- G4226A 1290 Infinity Autosampler
- G1316C 1290 Infinity Thermostatted Column Compartment
- G4212A 1290 Infinity Diode Array Detector
Conclusions
The combined high flow and high pressure capability of the system allows one to use high efficiency columns, producing rapid separations with remarkable resolution while conserving solvent over the use of 4.6 mm id columns. Impurity detection, due to high detector sensitivity and stability, is estimated to be ? 0.01%.
Agilent Technologies, Inc.
2850 Centerville Road, Wilmington, DE 19808
tel. (800)227-9770; fax (302)633-8901
Website:
Articles in this issue
over 15 years ago
Call for Application Notesalmost 16 years ago
Preparing for Pittcon 2010almost 16 years ago
The Analysis of Melamine in Milk Powder Using HILIC with UV Detectionalmost 16 years ago
Flexible Fused Silica Capillary: Impact of Aging on Strengthalmost 16 years ago
Analysis of Basic Pharmaceuticals and Peptides on a Multi-Mode ODS Columnalmost 16 years ago
Simplifying Peptide BioanalysisNewsletter
Join the global community of analytical scientists who trust LCGC for insights on the latest techniques, trends, and expert solutions in chromatography.
