The flexibility to use short sub-2 micron columns on UHPLC instruments at high flow-rates permits fast analysis with high resolution of complex samples. An Agilent ZORBAX Rapid Resolution High Definition (RRHD) Eclipse Plus C18 1.8 µm column separated seven different biocides in less than 1 minute with high resolution. A flow-rate of 1.7 mL/min was used on 2.1 × 50 mm column to achieve the rapid separation. The Agilent 1290 Infinity LC system was used as the column pressure just exceeded 1000 bar at this high flow-rate. A sample of hand sanitizer was found to contain 2-phenoxyethanol and methylparaben using the method.